Technical Discussion
N-550
101
Figure 44: Oxyhemoglobin Dissociation Curve
SatSeconds Alarm Management
The N-550 utilizes Nellcor’s
SatSeconds
alarm management technique.
SatSeconds
is a function of the software within the N-550. With the
SatSeconds
technique, upper and lower alarm limits are set in the same way as
traditional alarm management. The clinician also sets a
SatSeconds
limit that
allows monitoring of %SpO
2
below the selected lower alarm limit for a period
of time before an audible alarm sounds. Refer to the N-550 Operator's manual
for managing
SatSeconds
.
Reads Through Motion
The N-550 takes advantage of increased microprocessing power with
advanced mathematical algorithms.
O
XI
M
AX
advanced signal processing
allows the N-550 to read through challenging motion conditions to deliver
accurate saturation and pulse rate values. For a definition of motion, as
applicable to the N-550, contact Nellcor's Technical Services Department.
Summary of Contents for OxiMax N-550
Page 1: ...N 550 Pulse Oximeter Service Manual min min SpO2 N 550 SatSeconds SpO2 ...
Page 10: ...Blank Page ...
Page 38: ...Blank Page ...
Page 64: ...Blank Page ...
Page 81: ...Disassembly Guide N 550 71 Figure 39 Equipotential Ground Terminal Removed ...
Page 82: ...Blank Page ...
Page 86: ...Spare Parts 76 Figure 40 Exploded View ...
Page 90: ...Blank Page ...
Page 108: ...Blank Page ...
Page 120: ...Blank Page ...
Page 130: ...Figure 56 Main PCB Parts Locator Diagram 129 Technical Description N 550 ...
Page 131: ...Figure 57 Front Panel Parts Locator Diagram 131 N 550 Technical Description ...
Page 136: ...Blank Page ...
Page 137: ... ...