22
Illumination
Level
Light
Intensity
at 40 cm
distance
(
lux
)
Approx. time until end of use
Steady Illumination
Decreased
Illumination
Batteries
Empty
Green Lamp lit
Green and
Orange Lamp lit
Orange Lamp
starts to blink
Green Lamp
fades
High
14,000
4 hours
Orange Lamp lights with 30 min. left
Several minutes
Illuminator
does not light
Middle
9,500
7 hours
Orange Lamp lights with 35 min. left
Low
5,000
17 hours
Orange Lamp lights with 50 min. left
Note: The available illumination time becomes approximately 4, 7 and 17 hours at the “High”,
“Middle” and “Low” brightness levels, respectively. When the available illumination time
decreases and the remaining available time of use becomes 30, 35 and 50 minutes at
the “High”, “Middle” and “Low” brightness levels respectively, the orange lamp will light.
When the available illumination time elapses, the orange lamp will start blinking.
Charge the batteries immediately or replace with fully charged batteries. When the
batteries have been fully discharged, the green lamp will fade out.
5-5-3 Adjusting Illumination Field
The illumination field size can be changed continuously by turning the Diaphragm ring on the
Illuminator. The field size may be adjusted from 48 to 100mm approximately at a 40cm of
distance. By turning the Diaphragm ring clockwise, a wider illumination field size may be
achieved as shown in Fig. 5-5-4.
Fig. 5-5-3
Diaphragm Ring
Fig. 5-5-4
Diaphragm Ring