4.8.4 External Trigger Exposure Mode (Ext Edge 1, 2)
In external trigger mode (Trigger Dual Edge Mode), exposure time is set through
external trigger signals CC1 and CC2. The line period is set by the period of external
trigger signal CC1, and the exposure start is set by the rising edge of external trigger
signal CC1. The range of programmable exposure time, and the timing chart of the
exposure and the readout are shown below.
Exposure
Readout
①
①
②
Trigger
②
①
②
③
③
(CC1)
c
b
a
r
Trigger
(CC2)
Figure 4-8-4-1 External Trigger Exposure Mode (Ext Edge 1, 2)
The data of Exposure (1) is read out at Readout (1)
C in the above figure must not be less than 92.4
s
When using External Trigger Exposure Mode
(Trigger Dual Edge Mode)
, exposure time (
“a” in
the above figure) must not be less than 90.4
s.
Valid data 2098 pixel
Invalid data(0)
Invalid data(0)
XClk
YClk
(25M)
CST
VIDEO
LVAL
CC1
92.4
μ
s<
2098CK
100ns<
<2
μ
s
Exposure time>90.4
μ
s
CC2
100ns<
<2
μ
s