background image

necdisplay.com

NEC MultiSync® PA Series

Medical Applications

Creating a Clinical Review Preset

NEC MultiSync PA Series professional displays include a number of industry-leading features that make 

them ideal for a clinical review environment. Features includes a selectable factory preset DICOM GSDF 

setting for medical imaging and image uniformity compensation for optimum image quality. The built-in 

backlight sensor provides a feedback loop for brightness stability and consistency. Calibration is 

accomplished using 14-bit lookup tables (LUTs) that give you incredibly precise control over grayscale and 

color spaces.

The MultiSync PA Series is not shipped from the factory in the clinical review mode, however, through the 

use of NEC’s MultiProfiler™ software, the display can easily be set up for clinical imaging.

About Picture Mode Presets

MultiSync PA Series displays have five built-in Picture Mode presets that are available 

using the vertical hotkeys on the bezel, giving one-button access to up to five different 

display settings. The Picture Mode presets include customized color gamut, gamma, 

brightness, black level and white point settings. Any of the default picture modes on a 

MultiSync PA Series display can be easily customized using the free MultiProfiler tool. 

Creating a DICOM Preset

The Digital Imaging and Communications in Medicine (DICOM) standard defines a Grayscale Definition 

Function (GSDF) gamma curve optimized for diagnostic image viewing. The MultiSync PA Series includes a 

DICOM configuration preset that can be easily set in MultiProfiler. 

Before creating the preset, the uniformity should be adjusted to the desired 

level. The  Digital Uniformity Correction control adjusts the screen uniformity 

correction applied to the display in order to improve the screen luminance and 

color uniformity. This compensation is done at the expense of some screen 

brightness. A higher level of correction will give a more uniform screen, but the 

maximum brightness that can be achieved will be 

lower. Start the MultiProfiler tool and choose the 

Settings panel from the icons on the left. Using 

the slider, change the Digital Uniformity 

Correction to the desired level.

 

Next, select the Picture Mode panel. The Picture Mode panel controls the 

settings for the various Picture Modes that the monitor supports, as well as 

adjusting the Brightness and Black Level. The monitor supports five different 

configurable Picture Modes that can be quickly selected by clicking on the 

relevant buttons. Any picture mode can be customized by selecting the 

Picture Mode preset button (the display may flash while adjusting to the new 

picture mode) and click the “Edit…” button. 

The Color Configuration Wizard will appear with several predefined Picture 

Modes. For the clinical review mode, the DICOM preset should be selected. 

The wizard will then prompt for the desired white point. The default white 

point is 6500K (D65) and should be used, though Clear Base and Blue Base 

are options as well.

The next step is to adjust the color gamut 

for the display. There is no change required 

to the color gamut for DICOM GSDF, 

though several standard gamuts are available including sRGB and 

AdobeRGB (for wide gamut displays).

Finally, the brightness and black level can then 

be adjusted to the desired levels to achieve 

both the needed luminance and contrast ratios. 

Since several MultiSync PA Series displays have 

backlight sensors, the contrast ratio can be adjusted directly by changing the 

ratio of brightness and black level.

After completing these steps, a DICOM preset will 

be available in MultiProfiler as well as through the 

On Screen Display (OSD). To activate the OSD, 

use the vertical arrows on the bezel to display the 

list of Picture Modes. The same vertical arrows 

can be used to select the desired Picture Mode. The Picture Mode is 

automatically applied after being selected. 

Reviews: