Replenishment of the bacterial filter :
25
Bacterial filter is used in order to filter the air absorbed
from the environment which given to the patient with
the steam. Therefore, after a certain period of
use must be replaced with a new one.
10. CLEANING AND DISINFECTION
Jar and lid: Autoclave 121C-20mins recommended
Air and steam hose: After every use Autoclave 121C-20mins
recommended
Device body and the stand spare parts: Wipe with a cloth soaked in soapy
water and thoroughly wrung
11. TECHNICAL PROPERTIES
VOLTAGE
230 VAC ± % 10 , 50 Hz
POWER
40 W – 70 W ( WITH HEATER )
ELECTRICAL TYPE/CLASS
CLASS I , TYPE BF
OSCILLATOR FREQUENCY
1,68 MHz ± % 5
PARTICLE SIZE
0,5 – 6 µm
NEBULIZATION CAPACITY
0 – 5 ml / min ± % 10
AEROSOL TANK
MAX. 750 ml
WEIGHT
< 4 KG
AEROSOL TEMPERATURE
MAX . + 38 º C (WITH HEATER)
OPERATING TEMPERATURE
+ 10 ~ + 30 º C
OPERATING MODE
CONTINUOUS