A E - 1 5 0 s y s t e m – U s e r M a n u a l
43
21-
Acoustic channel 3
: Input BNC connector for acoustic sensor #3.
22-
HF partial discharge input
: Input connector for high frequency partial
discharge signals. Used to connect an HFCT clamp, a UHF antenna, a
flexible UHF antenna or another appropriate transducer (see
specifications).
23-
Charger port
: 12 Volts DC input for battery charging. The AE-150 can be
used with the battery charger connected. The battery automatically
charges when the charger is connected. Typical recharge time is 3 hours.
24-
Phase input
: Input connector for an ammeter clamp for phase
synchronization (optional) or another phase signal source (see channel
specifications).
25-
Acoustic channel 4
: Input BNC connector for acoustic sensor #4.
26-
Communication port:
Used for wired serial communication between the
AE-150 main unit and a computer. Also used to update the AE-150
firmware.