Voltage 
fluctuations/
flicker emissions
IEC 61000-3-3

Group 1

Class B

N

.

A

.

N

.

A

.

Guidance and manufacturer’s declaration–electromagnetic

emissions

Emissions            

Compliance     

Electromagnetic 

environment-guidance

RF emissions
CISPR 11

RF emissions
CISPR 11

Harmonic 
emissions
IEC 61000-3-2

Appendix 1 Guidance and Manufacturer Declaration Tables

The Model 

PG-IRT1603

 Infrared Thermometer is intended for use in 

the  electromagnetic  environment  specified  below.  The  customer or 
the  user  of  the  Model 

PG-IRT1603

  Infrared  Thermometer 

should assure that it is used in such an environment.

The Model 

PG-IRT1603

 Infrared 

Thermometer uses RF energy only 
for its internal function. Therefore, 
its RF emissions are very low and 
are not likely to cause any 
interference in nearby electronic 
equipment.

The Model 

PG-IRT1603

 Infrared 

Thermometer is used in home 
and it’s powered by DC 3V

Guidance and manufacturer’s declaration – electromagnetic 

immunity

The Model 

PG-IRT1603

 Infrared Thermometer is intended for use in 

the  electromagnetic  environment  specified  below.  The  customer or 
the  user  of  the  Model 

PG-IRT1603

  Infrared  Thermometer 

should assure that it is used in such an environment.

Immunity 

test

Electrostatic
discharge 
(ESD)IEC 
61000-4-2

IEC 60601

test level

±

8 kV contact

±

2

 

k

V

, ±

4

 

k

V

±

8 kV, ±15KV

 

air

Power 
frequency
(50/60 Hz)
magnetic 
field IEC 
61000-4-8

30 A/m

50/60Hz

NOTE    U  is the a.c. mains voltage prior to application of the test level

T

Compliance 

level

Electromagnetic 

environment-guidance

±

8 kV contact

±

2

 

k

V

, ±

4

 

k

V

±

8 kV, ±15

 

K

air

Floors should be wood, 
concrete or ceramic tile. 
If floors are covered with 
synthetic material, the 
relative humidity should 
be at least 30 %.

30 A/m

50/60Hz

Power frequency 
magnetic fields should 
be at levels characteristic 
of a typical location in a 
typical commercial or 
hospital environment.

Specifications of the product

Product name:

  Infrared 

Forehead/

Ear Thermometer

 

Model number: 

PG-IRT1603

Product appearance dimensions: 

31×175×72mm

 

Product weight:

 about 77g (

without

 battery)

Measuring range: 

93.2

°

F

-

109

.

4

°

F.

Object temperature:

 

32°F-199.7°F

 (

0°C- 93.2°C

)

 

Resolution ratio:

 0.1°C

/

°F

Measurement Location:

 laboratory

Accuracy:

 

(

95.0°F~107.9°F

)

±0.4°F

,(

35.0°C~42.0°C

)

±0.2°C, 

other temperature

±

0.3°C

.

Operation

 

temperature: 

50.0°F~104.0°F 

(

10.0°C~40.0°C

)

relative maximum humidity 15%RH

~

93%RH

 

Atmospheric pressure:

70kPa~106kPa

Transportation/storage temperature:

 -13°F~131°F

(

-25°C

~55°C

)

relative maximum humidity 0%RH

~

93%RH

 

Atmospheric pressure: 

50kPa~106kPa

Display screen:

 LCD display screen, 4 bit numbers and 

special icons.

Sound: 

when you turn on the product and 

are 

ready to 

measure,  a short beep will be heard.  

The measurement is finished with a long beep.

 

System error or fault: 

beeps quickly three

 times.

 

Fever alert: 

beeps quickly

 ten times with urgency.

1. Main body
2. Product manual

Memory:

 

record nine temperature settings in memory 

mode

.

Automatically shut down:

 if no operation for 30 seconds,  

the device

 will shutdown automatically.

Battery:

 two 1.5V AAA  batteries 

(alkaline batteries are recommended).

Period of use: 

five years

1603 Forehead mode

Clinical bias, Dcb: 0.078
Limits of Agreement, LA

0.243 

Clinical Repeatability, 

σ

r

 

0

  .

069

R

eference body site: forehead

Measuring site: forehead

Packing parts list

: 

Summary of Contents for PG-IRT1603

Page 1: ...mometer Thanks for buying and using this product please read this manual carefully before use User manual The USB Lighter Company LLC Address 10940 Wilshire Blvd STE 2010 Los Angeles CA 90024 949 682 8817 Brand NCT Date 06 16 2020 Rev A 0 ...

Page 2: ...ease do not use this product for purposes other than body temperature measurement Do not use mobile phones close to this product Please do not use equipment that generates electromagnetic fields near the product Please do not disassemble or repair this device by yourself Please do not bend or stretch the device Please do not impact or drop this product Caution please refer to attached file Legend ...

Page 3: ...outh When the infrared sensor is damaged or dirty it may cause abnormal measurement results 3 If there is a temperature difference between the storage site and the measurement site please place it at room temperature measurement site for about 30 minutes before next measurement May result in incorrect measurement results 4 Please keep this product out of the reach of children Children can potentia...

Page 4: ...idue of this product at the end of the service life according to local laws and regulations Common sense about body temperature The comparison of different measurement methods The measured values are different if we use different measurement methods The WHO provides normal human body temperature reference values please see below table about the specific temperature difference The changes in human ...

Page 5: ...s low LCD shows sign and battery mark is always on which means you should replace the battery Warm prompts If storing the product for an extended period please remove the battery to assure longevity Battery fluids may harm the product and pollute our environment It is recommended to use alkaline batteries The way to handle a scrapped battery should match the requirements of local country governmen...

Page 6: ... Measurement result Note If you did not hear the beep sound this means that the temperature measurement has not yet been completed Please do not remove the thermometer probe from forehead at this time If you have disabled the sound indication it will have no sound indication 3 Object pattern measurement The installation of the head cover of forehead temperature has no influence to object pattern m...

Page 7: ...H When you operate this product the environment temperature cannot be greater than 104 F Er L This product has a lower operating temperature of 50 F When the ambient temperature exceeds this temperature point the LCD screen will display an error message Er L When you operate this product the environment temperature can not be lower than 50 F The measurement temperature is low The measurement posit...

Page 8: ...kV 4 kV 8 kV 15 KV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 30A m 50 60Hz Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Specifications of the product Product name Infrared Forehead Ear Thermometer Model number...

Page 9: ...odel PG IRT1603 Infrared Thermometer is used exceeds the applicable RF compliance level above the Model PG IRT1603 Infrared Thermometer should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the Model PG IRT1603 Infrared Thermometer d Over the frequency range 150 kHz to 80 MHz field strengths should ...

Page 10: ...ed Maximum output of transmitter W Separation distance according to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 7 GHz P V d 5 3 1 P E d 5 3 1 P E d 7 1 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated usin...

Reviews: