13
14
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Group 1
Class B
N
.
A
.
N
.
A
.
Guidance and manufacturer’s declaration–electromagnetic
emissions
Emissions
Compliance
Electromagnetic
environment-guidance
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Appendix 1 Guidance and Manufacturer Declaration Tables
The Model
PG-IRT1603
Infrared Thermometer is intended for use in
the electromagnetic environment specified below. The customer or
the user of the Model
PG-IRT1603
Infrared Thermometer
should assure that it is used in such an environment.
The Model
PG-IRT1603
Infrared
Thermometer uses RF energy only
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
The Model
PG-IRT1603
Infrared
Thermometer is used in home
and it’s powered by DC 3V
Guidance and manufacturer’s declaration – electromagnetic
immunity
The Model
PG-IRT1603
Infrared Thermometer is intended for use in
the electromagnetic environment specified below. The customer or
the user of the Model
PG-IRT1603
Infrared Thermometer
should assure that it is used in such an environment.
Immunity
test
Electrostatic
discharge
(ESD)IEC
61000-4-2
IEC 60601
test level
±
8 kV contact
±
2
k
V
, ±
4
k
V
,
±
8 kV, ±15KV
air
Power
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
30 A/m
,
50/60Hz
NOTE U is the a.c. mains voltage prior to application of the test level
T
Compliance
level
Electromagnetic
environment-guidance
±
8 kV contact
±
2
k
V
, ±
4
k
V
,
±
8 kV, ±15
K
V
air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
30 A/m
,
50/60Hz
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
Specifications of the product
Product name:
Infrared
Forehead/
Ear Thermometer
Model number:
PG-IRT1603
Product appearance dimensions:
31×175×72mm
Product weight:
about 77g (
without
battery)
Measuring range:
93.2
°
F
-
109
.
4
°
F.
Object temperature:
32°F-199.7°F
(
0°C- 93.2°C
)
Resolution ratio:
0.1°C
/
°F
Measurement Location:
laboratory
Accuracy:
(
95.0°F~107.9°F
)
±0.4°F
,(
35.0°C~42.0°C
)
±0.2°C,
other temperature
±
0.3°C
.
Operation
temperature:
50.0°F~104.0°F
(
10.0°C~40.0°C
)
,
relative maximum humidity 15%RH
~
93%RH
Atmospheric pressure:
70kPa~106kPa
Transportation/storage temperature:
-13°F~131°F
(
-25°C
~55°C
)
,
relative maximum humidity 0%RH
~
93%RH
Atmospheric pressure:
50kPa~106kPa
Display screen:
LCD display screen, 4 bit numbers and
special icons.
Sound:
when you turn on the product and
are
ready to
measure, a short beep will be heard.
The measurement is finished with a long beep.
System error or fault:
beeps quickly three
times.
Fever alert:
beeps quickly
ten times with urgency.
1. Main body
2. Product manual
Memory:
record nine temperature settings in memory
mode
.
Automatically shut down:
if no operation for 30 seconds,
the device
will shutdown automatically.
Battery:
two 1.5V AAA batteries
(alkaline batteries are recommended).
Period of use:
five years
1603 Forehead mode
:
Clinical bias, Dcb: 0.078
Limits of Agreement, LA
:
0.243
Clinical Repeatability,
σ
r
:
0
.
069
R
eference body site: forehead
Measuring site: forehead
Packing parts list
: