CS400
System
Hardware Manual, Rev. 1.0, Sep 2021
24
Figure 4-4: Opened System Connection Box
Installation
The system connection box can be mounted up to 7 meters (23 ft) away from the scanner. However, to simplify
aiming the scanner at the kiln it is the best that the connection box is mounted as close to the scanner as possible.
The cables between the linescanner and the connection box needs to be protected from mechanical damage.
Mount the connection box in a convenient location. The cables W1, W3, and W18 are factory preinstalled. Insert
the cable W20 using the grommet plate taking care to select the correct grommet size for the cable diameter.
Connect the cables for the accessories with the connectors of the line-up terminal as described in section
8.4
After double-checking all connections, switch the power on. The 24V-LED indicates the ON/OFF status (see
terminal pin 8 in the system connection box). Check the trigger signal coming from the position indicator (see
terminal pin 40 in the system connection box).
Pins 17 and 18 of the terminal in the system connection box connect to the internal alarm relay of the scanner.
The contacts are potential free, the maximum load is 30 V / 1 A. The configuration of the alarm relay is supported
by the CS400 system software.
Figure 4-5: Connection to the Internal Alarm Relay of the Scanner
Mains
Socket
Fuse
6 A
Power
Supply
Serial/Ethernet
Converter
(Accessory)
Fiber Optic /
RJ45 Ethernet
Converter
Heater
(Accessory)
Pin 18
Pin 17