Pre-Installation
Cable Requirements
19
3
3.5 Cable Requirements
The following cables are necessary for standard installations, see also system drawings given in section
8.2
•
W1
power supply cable for scanner (preinstalled with system connection box)
•
W3
trigger/alarm cable for linescanner (preinstalled with system connection box)
•
W8
from the junction box of the position indicator to the system connection box (standard installation) or
to TSM remote control box (when used with accessory Tire Slip Monitoring).
•
W9
from the TSM remote control box to the system connection box. This cable is used for data
communication, power supply, and trigger pulse.
•
W14
from the MI connection box to the system connection box and between MI connection boxes when
more than one shadow sensor is installed. This cable is used for data communication and power supply
to the pyrometer.
•
W15
from the Endurance connection box to the system connection box. This cable is used for data
communication and power supply to the pyrometer.
•
W16
MI sensing head cable.
•
W17
Endurance sensing head cable.
•
W18
Ethernet communication cable for scanner (preinstalled with system connection box)
•
W20
Fiber optic cable from system connection box to the control room and to the second scanner (if
used). Fiber optic cable to be provided by customer.
•
W22
Ethernet cable from fiber optic converter (control room) to the PC.
•
W23
from the position indicator head to the TSM junction box.
•
W24
Power supply cable 100 to 240 VAC.
•
W25
Fiber optic cable from the first scanner to the second scanner (if used). Fiber optic cable to be
provided by customer.
Note
All copper cables must be shielded! The wires from W9, W14, and W15 must be a twisted pair! Local building
codes should be observed when selecting cables!
Table 3-1: Required Cables
Cable
Distance
Cable features
Supplied from …
Example / Remarks
W1, W3
7.5 m (25 ft.)
preinstalled
Manufacturer
W8
350 m (380 yd)
3 x 0.25 mm²,
24 AWG, 3 conductors, shielded
Customer
(N)YLHCY-J 3 × 0.25 mm²
Manhattan/CDT, P/N M13233
W9
350 m (380 yd)
2 x 2 x 0.25 mm²,
24 AWG, 3x twisted pair, shielded
Customer
LifYCY 2 × 2 × 0.25 mm²
W14
350 m (380 yd)
2 x 2 x 0.5 mm²,
20 AWG, 2 x twisted pair, shielded
Customer
max. 5 pyrometers
350 m (380 yd)
2 x 2 x 1.5 mm²,
16 AWG, 2 x twisted pair, shielded
Customer
max. 14 pyrometers
W15
350 m (380 yd)
2 x 2 x 1.5 mm²,
16 AWG, 2x twisted pair, shielded
Customer
W16
8 m (26 ft)
preinstalled
Manufacturer
W17
15 m (49 ft.)
preinstalled
Manufacturer