Chapter 11 METER SETTING
11.5 Compact Flash Setting
(
F-53, 54, 55
)
146
HORIBA
WRITE OPERATOR INFORMATION
Information such as Operator names or currently set configuration is copied onto the compact
flash card.
Personal data can be written only for one person per each CF card.
1.
Press the ENTER key.
Personal data such as Operator name, meter configuration items, each setup item is
stored into the compact flash card.
FORMAT
The compact flash card is formatted.
Formatting will delete all the data previously stored.
1.
Press ENTER key.
2.
Press F4 (CONFIRM).
The formatting starts.
Since the formatting uses a part of memory, the empty capacity displayed will be smaller than
that mentioned in the compact flash card even though no data is stored.
11.5.4 Data acquisition to PC
To transfer the data from the compact flash card, the following tools are necessary:
●
PC
●
A compact flash card reader (those commercially available)
●
Software
The software is available by downloading from HORIBA’s website after user registration.
Fill in the User Registration Card separately attached to this manual to complete the
registration.
Summary of Contents for F-52
Page 2: ......
Page 13: ...TABLE OF CONTENTS X HORIBA...
Page 31: ...Chapter 1 Overview 1 9 Security Function 18 HORIBA...
Page 41: ...Chapter 2 BASIC OPERATIONS 2 5 Navigation Function 28 HORIBA...
Page 103: ...Chapter 5 Ion Measurement F 53 55 5 5 Check Display 90 HORIBA...
Page 125: ...Chapter 8 RESISTIVITY MEASUREMENT F 54 55 8 2 Setting Configuration 112 HORIBA...
Page 165: ...Chapter 11 METER SETTING 11 8 FDA Part 11 Function Audit Trail F 55 152 HORIBA...
Page 187: ...Chapter 13 ANALOG OUTPUT 13 3 Alarm Output 174 HORIBA...
Page 227: ...Chapter 15 Maintenance and troubleshooting 15 2 Troubleshooting 214 HORIBA...