3 Functions
3.4 Setting modes
80
HORIBA
3.4.4 Temperature compensation setting
1.
Press the SET key in the Measurement mode to
enter the Temperature Compensation Setting
mode.
2.
Pressing the ENTER key toggles between MTC
and ATC settings.
ATC
Automatic temperature compensation (when using a
temperature sensor of the electrode)
ATC is displayed.
When a temperature sensor is connected, the current
temperature is automatically displayed.
(When no temperature sensor is connected, the display
shows 25
°
C.)
MTC
Manual temperature compensation (when an electrode
temperature sensor is not being used and the
temperature of the solution is known before hand)
MTC is displayed.
Set the temperature using the
▲
and
▼
keys.
Setting range: 0.0 to 100.0
°
C