Single Use Pre-

Sized XactTrace™ Belt System User Instructions 

System Components 

The system consists of a belt and a snap sensor with a cable that 
connects directly to an input on the recording device. The system 
converts changes in inductance to a digital signal that provides both 
qualitative and quantitative information on respiratory effort. An optional 
RIP Processor can be purchased to interface the locks with most PSG 
systems that accept the 1.5mm female touchproof connectors. 

 

 

Belt 

 

Sensor 

 

Fitting the XactTrace Belts 

The belts are intended to be worn over nightclothes and should fit the 
patient snugly without being uncomfortably tight. Avoid all unnecessary 
contact with moisture when using the Single Use Pre-Sized XactTrace 
belt system. 

 

Note:

 Do not use two thorax sensors or two abdomen 

sensors in the same recording. Using two sensors of the 
same type will cause interference between the sensors and 
could result in poor signal quality. 

Summary of Contents for XactTrace

Page 1: ...and attaching the belts storage and cleaning Intended Use XactTrace is intended to measure respiratory effort to assist in the diagnosis of sleep disorders or sleep related respiratory disorders The r...

Page 2: ...other words in the presence of flammable liquids such as an anesthetic mixture with air or with oxygen or nitrous oxide Connect the XactTrace sensors only to an input that is electrically isolated fr...

Page 3: ...an be purchased to interface the locks with most PSG systems that accept the 1 5mm female touchproof connectors Belt Sensor Fitting the XactTrace Belts The belts are intended to be worn over nightclot...

Page 4: ...gly without being uncomfortably tight Measure the patient s circumference and use the table below as a guideline The belts are single use XactTrace Belt Sizes Full Size Reduced Size cm Inches cm Inche...

Page 5: ...ector as shown in the figure below The belt should fit snugly without being uncomfortably tight 2 Prepare the abdomen belt in the same manner only this time fit the belt around the patients stomach an...

Page 6: ...sor module labels are color coded for easy identification and connection 5 Plug the thorax and the abdomen module outputs into the appropriate bipolar inputs on your polygraph The red touch proof plug...

Page 7: ...Fitting the XactTrace Belts Material Elastic belt W 25 WHITE TERYLENE WITH WIRE UL1571 28 WHITE Cable Length 16 inches 41 cm Thorax 26 inches 66 cm Abdomen or 78 inches 200 cm Thorax Abdomen Material...

Page 8: ...heric pressures from 7 3 psi to 29 psi Output Specifications RIP Processor Output Signal Maximum signal amplitude 2mV Frequency Range 0 2 to 3 Hz Sensitivity Approximately 50 V mm Storage Proper stora...

Page 9: ...fects in materials and workmanship for 6 months from the date purchased The sole liability of Embla and our distributors is limited to replacement or repair of the product at the option of Embla with...

Page 10: ...nditions of sale of Natus Neurology Incorporated or its distributors Natus Neurology Incorporated makes no representations or warranties of any kind whatsoever with respect to this document Natus Neur...