Use the cleaning wire to clean the sound channels of the probe tip from the
rear.
Note
•
Remember to clean the cleaning wire when it protrudes from the probe
tip.
When you have finished using the cleaning wire, then remember to disinfect the clean-
ing wire in accordance with local procedures.
Disinfecting procedures for the probe tip
The probe tip material is highly resistant to a wide range of temperature and chemical influences.
In order to disinfect the probe tip, perform the following:
•
After cleaning, immerse the probe tip in a bath of 70-90% ethyl or isopropyl alcohol for 10-30 minutes contact time.
When you have disinfected the probe tip, rinse it thoroughly in regular water.
If your established infection control procedure dictates autoclaving, make sure that the probe tip has not been deformed
by the autoclaving process.
Make sure that the sound channels are completely dry before you fit the tip back onto
the probe body. If needed, use a spare probe tip.
10.1.2
Cleaning the probe body
Important
•
For periodical cleaning of the probe body, contact your authorized service department.
Caution
•
The probe body contains sensitive components. Never clean the sound apertures in the probe body mech-
anically or with liquids. Doing so may cause damage to the probe.
Caution
•
Even the slightest amount of moisture may dissolve any residual cerumen and contaminate the sensitive
parts in the body of the probe.
–
Use a moist tissue with alcohol for regular surface cleaning.
Caution
•
No part of the ear probe should be subjected to ultrasonic cleaning solutions or machines.
Otometrics - MADSEN Alpha OAE+
63
10 Maintenance and Cleaning