
8.2.1
Performing the probe test
1. Insert the probe tip without eartip in the test cavity.
The test starts automatically.
Note
•
The probe test does not start automatically e.g. when a patient is being
created.
If the probe functions correctly, the message
Probe OK
appears.
If the probe does not function correctly, the message
Probe failed
appears.
You will also see a list of possible errors.
2. Press the
OK
button to confirm the test result.
If needed, repeat the test.
Otometrics - MADSEN Alpha OAE+
51
8 Alpha OAE+ settings