For a detailed description of traditional FM testing, see the AURICAL HIT Reference Manual.
5.10
How to perform a standard test
The procedure
1. Launch the fitting software for the hearing instrument so that you can control its parameters.
2. Launch OTOsuite and select
HIT
in the
Navigation
panel.
3. Open the
Test Selector
and select the
ANSI
or
IEC
special test.
4. If OTOsuite is used without Noah, you can fill out the
Hearing Instrument
fields in the lower right corner of the
Standard Tests
screen. These fields are filled in automatically when you use OTOsuite with Noah together with the fit-
ting software for the hearing instrument.
5. Position the hearing instrument so that it is ready for testing, and switch it on.
6. If you wish to measure the
Battery Consumption
, make sure that you connect the battery simulator.
7. Close the lid.
8. If needed, click the arrow buttons in the
Measurements
table to include the individual tests you wish to perform.
9. Click the
Start
button in the top left corner of the
Measurements
table.
This will start a sequence of selected tests.
10. Make sure that you follow the on-screen instructions.
11. If you wish to redo an individual test, click on the
Start
button next to the test.
Otometrics - AURICAL HIT
15
5 Testing hearing instruments