6-2
OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Repair Procedures
Replacing the Main and Transformer Fuses
If the device is plugged in but won’t turn on, the fuse may have to be replaced.
Required Items:
Screwdriver
To replace the fuses:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Remove the upper white door seal and loosen—but do not remove—screws just
inside the Drier door (behind the door stop) (see Figure 4.2 on page 4-3).
3
Pull out and remove the front cover (see Figure 4.3 on page 4-3).
4
Locate, remove, and replace the fuse(s).
5
Position the front cover, tighten the screws, and push the upper white door seal
into place.
6
Reconnect the power cord to the electrical outlet.
7
Test the device for proper operation (see page 2-4).
CAUTION
Replace the fuse only with a new fuse of the same type and value.
WARNING
Electrical shock hazard when the front panel is removed.