Performing a Cortical Stimulation Study
Using the Nicolet Cortical Stimulator to set the Cortical
Stimulator parameters
NOTE:
Use the
Selector
knob to choose the desired parameter from the display.
1. Press the
Parameter
button
for the parameter you want to set up.
2. Turn the
Selector
knob
to scroll through the options.
3. Press the
Selector
knob
or
Parameter
button to select the desired option.
IMPORTANT: The stimulus output is stopped automatically if the current
delivered exceeds the requested current level by 20% or greater. The Nicolet
Cortical Stimulator must be turned off and then turned back on to clear this
condition.
If you are currently stimulating the patient, press the Stop button on the
Stimulus Control Unit (SCU) before changing any parameters.
Nico
let
Cor
tical Stimula
tor
Status
A B C...
Hz
Se
tup
Pulse F
requenc
y
Pulse Dur
ation
Elec
trode -
Elec
trode +
Train
Du
ration
Deliv
ered Stim
Set Stim
Check Stim
Ictal
Disrupt
Channel M
ark
Star
t
Stop
Selector knob
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...