User and Service Manual
Natus® Brain Monitor &
Embla Dx Series™
15
Perform the recommended maintenance. Refer to the
further details.
Do not immerse the amplifier or any of its components in water or other fluid.
Proper use of this device depends on the careful reading of all instructions and labels that come
with or on the system. Inaccurate measurements may be caused by incorrect application or use.
When replacing the fuse for the Brain Monitor & Embla Dx Series Amplifiers, it must be replaced
by a fuse with the same type and rating as the original fuse. Replacement
purchased from Natus directly.
The Brain Monitor & Embla Dx Series Amplifiers are compatible with NeuroWorks / SleepWorks
v9.0 and newer.
The amplifier base unit is classified as an IPX0
– ordinary degree of protection against ingress
of water according to IEC 60529.
The amplifier Breakout is classified as body worn and has an IPX1 rating while inside the pouch
(-61) and meets IEC 60601-2-26 spillage requirement without the pouch.
The system is classified as a class I device according to IEC 60601-1.
WARNING:
Third-party software installed on the acquisition computer may interfere with the
operation of the Natus software. Please consult Natus Technical Support before installing third-
party software on the computer.
WARNING:
No modification of this equipment is allowed.
Electrostatic Discharge (ESD) Precautions:
Electrostatic Discharge (ESD) Precaution:
Be sure to take the appropriate
Electrostatic Discharge (ESD) precautions. Disconnect the cables before moving,
cabling, or performing any set up procedures. Connectors marked with the ESD
protection symbol should not be touched. For detailed handling procedures, refer to
the
Electrostatic Discharge (ESD) Handling Procedures and Warnings
Turn off all system power and disconnect the power cord from the system and the wall before
attempting to clean the unit. The Brain Monitor & Embla Dx Series Amplifiers can be wiped
clean with a soft, damp cloth using non-conductive distilled water, electrically non-conductive
inert surfactants or a Natus approved cold sterilizing agent. It is important to dry off the unit
quickly. Avoid letting liquid seep into any of the internal electronics of the system. Do not use
any abrasive cleaner on the system.
Device accessories may include several kinds of disposable, sterile needle electrodes. These
needles are labeled as
STERILE
and the method of sterilization is documented on the
packaging. These electrodes should not be used if the sterile packaging has been tampered
with.