MADSEN Alpha OAE+ Screener User Manual
39
8
Alpha OAE+ settings
You can configure your Alpha OAE+ from the Alpha OAE+
Settings
menu.
8.1
Settings menu
From
Settings
(
), you can change Alpha OAE+ settings.
8.2
Probe Test
You should test the probe daily to ensure that it functions correctly. You must test the probe before patient testing.
Note: Make sure that the probe tip has been cleaned and disinfected before you insert it in the test cavity. See Cleaning the built-in test cavity
You can start the probe test in two ways:
Inserting the probe in the Alpha OAE+ test cavity (see
-or-
Go to Settings (
) to restart the test or to test the probe using an external test cavity.