Introduction
July 14, 2016
1-9
Cleaning recommendations
ErgoJust cart
All paints and plastic components will withstand cleaning by most commonly used, diluted, non-
abrasive solutions such as quaternary ammonia compounds, ammonia enzyme cleaners, bleach or
alcohol solutions. It is recommended that any cleaning solution be tested on a small, inconspicuous
area to ensure the surface is not harmed.
•
Pen and permanent and dry erase markers can be removed with 91% isopropyl alcohol and a soft
cloth.
•
Iodine stains can be removed with commonly used cleaners and a soft cloth.
•
Never
use steel wool or other abrasive materials that will damage the surface finish.
System components
Cleaning consists of removing all dust from the exterior surface of the system with a soft brush or
cloth. Use a brush to dislodge any dirt on or around the connectors and panel edges. Remove any
dirt with a soft cloth, slightly dampened with a mild detergent solution or cold sterilization agent.
For Natus amplifier devices, refer to the corresponding user and service manual.
Turn OFF system power before cleaning. Prevent detergent solution or cold
disinfecting agents from seeping into the electronics of the instrument. Be especially careful
around controls, connectors and edge panels. Do not use abrasive cleaners.
Preventive maintenance
Electrical safety testing is recommended. It is recommended a schedule be established for these purposes, with
at least an annual cleaning and safety testing. This system does not require calibration unless otherwise stated.
Preventative maintenance does not require access to the interior of the instrument and may be performed by
the user. For this device, preventative maintenance consists of periodically cleaning and inspecting the
exterior of the instrument.
It is recommended that all repairs be performed by a qualified Natus Neurology Incorporated service
representative only. You have the sole responsibility for any malfunctions resulting from improper
maintenance or repair by anyone other than an authorized Natus Neurology Incorporated representative.
If the system is not functioning properly, do not operate it until all necessary repairs are made and unit is
tested for proper functioning in accordance with Natus Neurology Incorporated published specifications. It is
recommended that all repairs be performed by a qualified service representative only.
Switch off all power to the system before attempting any service, maintenance, or preventive maintenance.
If the system or any component is repaired, it is recommended that all system functionality be checked and an
electrical safety test be performed prior to resuming use.