Sterilization cycle parameters for the fiber optic cable
EtO Sterilization Cycle
Parameters for Fiber Optic
Cable
Sterrad
®
100S Cycle Parameters for Fiber Optic Cable
Temperature
55 ±3 ºC
Pre-Conditioning
Pre-Plasma
10 minutes
Exposure Time
60 minutes
Phase 1 - Exposure 1
Injection
Diffusion
Plasma
6 minutes
2 minutes
2 minutes
Humidity
50-80% RH
Phase 2 - Exposure 2
Injection
Diffusion
Plasma
2 minutes
2 minutes
2 minutes
Aeration at
55ºC
> 8 hours
Injection H2O2 Volume (59% nominal
concentration by weight)
1800
μl
±50
μl
Concentration
725 ±25 mg/L
Following Sterilization
Allow cables to cool down to ambient temperature before starting the monitoring process. Before each use, visually
inspect the sterilized cables for any possible defects. Do not use defective cables.
Sterilizing the Camino
Flex Extension Cable
The Flex Extension Cable (REF # FLEXEXT) is provided non-sterile.
Caution
A sterile Flex Extension Cable is required during implantation of the catheter. Prior to sterilization, clean this cable using the
guidelines listed on Chapter 7.
Limits on Sterilization
The Flex Extension Cable may be sterilized thirty (30) times. Do not use this cable after reaching the sterilization
limit. For tracking purposes, use the reprocessing record sheet (doc # 026426) supplied with your cable.
Sterilization Parameters
The Flex Extension Cable supports steam sterilization only. Prior to sterilization, the cables should be packaged
individually in a self-sealing sterilization pouch with dimensions 7.5" x 13" or larger.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...