Cleaning Guidelines
Use the following guidelines when cleaning the Camino ICP Monitor and each of its system components that
are listed in the box below:
•
External surface
•
Touch screen
•
Natus Camino Fiber Optic Catheter Cable
•
Natus Camino Flex Extension Cable
•
Cable Strap
Note that each of these components should be cleaned immediately after contamination.
Warning
Only use the cleaning agents listed in this section for cleaning and disinfecting the Camino ICP Monitor system. Using
solvents or cleaning agents not listed in the cleaning guidelines may damage the plastic exterior of the Camino ICP
Monitor.
Warning
Do not autoclave or immerse the Camino ICP Monitor in liquid as damage may occur. If the monitor is exposed to liquids,
turn off the unit, remove the AC power adapter, dry the unit thoroughly, and send to biomed staff for evaluation before
reapplying power.
Caution
The Fiber Optic Catheter Cable and Flex Extension Cable should not be cleaned using automated cleaning methods.