91
Conditions That Reset Trend Data During Recording
Depending on the catheter type being used to record the patient’s Mean ICP trend data, the following conditions
will result in trend data being reset for that patient:
Catheter
Conditions
Natus Camino Fiber Optic (110-4
Series)
•
Monitor is turned off and then turned back on.
•
Battery discharges completely, resulting in the monitoring being
turned off and then turned back on.
•
The current ICP catheter is disconnected from the monitor and
then re-connected.
•
The current temperature catheter is disconnected from the monitor
and then re-connected.
•
The current catheter is replaced by a different catheter.
Natus Camino Flex
•
The current catheter is replaced by a different catheter.
Warning
Replacing a catheter will reset the trend data. Please extract any data that you wish to retain prior to replacing the catheter.