86
Responding to Battery Failure Alarm
If the battery control system inside the ICP Camino Monitor fails during the use of the monitor, the monitor will no
longer be able to report the amount of remaining charge that battery has. If this problem occurs, the monitor will
activate a low priority technical alarm that will:
•
Sound an alarm containing a burst of one single pulse every 16 seconds
•
Battery power symbol on the touch screen will appear crossed-out.
•
Displays the following message on the status bar:
"
Battery Failure"
If this battery failure alarm activates, do not run the monitor on battery power. If necessary, plug the monitor into
an AC outlet to maintain power. After patient treatment, try the following:
•
Turn off the monitor and then turn it back on.
•
Replace the battery.
If this alarm condition persists, contact Natus Neuro for service.
Silencing the Battery Failure Alarm Temporarily
Press the yellow alarm symbol on the touch screen. This will silence the alarm for 3 minutes. If the monitor
remains on for 3 minutes while there is still a reported battery failure, the alarm will reactivate.