82
Responding to Temperature Sensor Failure Alarm
If the temperature catheter fails during use, the monitor will activate a medium priority alarm that will:
•
Sound an alarm containing a burst of three pulses
•
Display the following error message on status bar: "
Temperature sensor failure"
•
Remove the ICP and temperature values from the screen
To resolve this problem, disconnect the temperature cable. To resume monitoring temperature, the entire
catheter must be replaced.
Silencing the Catheter Failure Alarm Temporarily
Press the flashing alarm symbol on the touch screen. This will silence the alarm for 3 minutes. If the catheter is not
disconnected within 3 minutes, the alarm will reactivate.
Responding to Low Battery Alarm
The Camino ICP Monitor includes a replaceable 14.4 V lithium ion battery that supplies power to the monitor for
at least 1.5 hours. The battery is intended to power the monitor during patient transport only; use AC power
whenever possible.
15 Minutes or Less of Remaining Power
If the battery power level drops to 15 minutes or less of remaining charge, the Camino ICP Monitor will display the
following message on status bar: "
Battery is low"
If this occurs, plug in the monitor to an AC power outlet. This
warning message will not disappear until the monitor is connected to AC power.
Summary of Contents for Camino ICP
Page 2: ...2...
Page 12: ...12 Blank page...
Page 16: ...16 Blank page...
Page 26: ...26 About the Left Panel Theleftpanel doesnotcontain anyusableconnector ports or buttons...
Page 56: ...56 Blank page...
Page 74: ...74 Blank page...
Page 92: ...92 Blank page...
Page 100: ...100 Blank page...
Page 110: ...110 Blank page...
Page 142: ...142 Blank page...
Page 146: ...146 Blank page...