Chapter 3 Maintenance
© National Instruments Corporation
3-5
VXI-1500 User Manual
Table 3-1.
Troubleshooting
Possible Cause
What to Do
VXI-1500 mainframe not
connected to power source.
Make sure that the VXI-1500 is connected
to a live electrical outlet. Try operating
another piece of equipment from this
outlet.
Power switch is not
switched on.
Set the front power switch to the On (1)
position.
Remote inhibit input on the
rear panel of the mainframe
is active.
De-activate the remote inhibit control
provided by your system.
Circuit breaker is tripped.
Reset the circuit breaker. Refer to the
Resetting the AC Mains Circuit Breaker
section in this chapter.
Power supply failure.
Contact National Instruments.