4. Power on the chassis.
Related Information
Refer to the NI Switches Help for more information about the connectors on the NI
50-Pin Rear Connector
You must determine the number of pins on the rear connector of your NI SCXI switch module
to choose the appropriate set of steps to cable the switch controller to the switch module.
Refer to the following figure and complete the following steps to cable the switch controller to
your NI SCXI switch module with a 50-pin rear connector for the NI 4021/407x or
NI PXI/PCI-4065.
Figure 5.
Cabling an NI 4021/407X or an NI PXI/PCI-4065 to a 50-Pin Rear Connector Switch
Module
1
2
4
5
3
1. NI SCXI Switch Module with a 50-Pin Rear
Connector
2. NI SCXI-1000/1001 or NI PXI-1010/1050 Chassis
3. NI SCXI-1362 Cable Adapter
4. NI Switch Controller
5. SH9MD-AUX Cable
1. Attach the NI SCXI-1362 cable adapter to the rear of your NI SCXI switch module.
2. Screw the cable adapter to the chassis mounting rail.
3. Connect one end of the SH9MD-AUX cable to the NI SCXI-1362 cable adapter and the
other end to the NI 407x.
NI Switches Getting Started Guide
| © National Instruments | 11
Summary of Contents for SCXI-1175
Page 1: ...SCXI 1175 ...