CALIBRATION PROCEDURE

PXIe-4309

32 Ch (8 ADC), 2 MS/s, 18 - 28 bit, Flexible Resolution PXI Analog Input 
Module

This document contains the verification and adjustment procedure for the PXIe-4309 module.

Note

To maintain forced air cooling in the PXI Express system, refer to the 

Maintain Forced-Air Cooling Note to Users

.

Contents

Software.................................................................................................................................... 1
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 4
Initial Setup............................................................................................................................... 4

Self Calibration................................................................................................................. 4

Verification............................................................................................................................... 5

DC Offset Accuracy ......................................................................................................... 6
DC Voltage Accuracy....................................................................................................... 9

Adjustment................................................................................................................................ 13
EEPROM Update ..................................................................................................................... 16
Reverification ........................................................................................................................... 17
Specifications............................................................................................................................ 17
Worldwide Support and Services ............................................................................................. 17

Software

Calibrating the PXIe-4309 requires the installation of NI-DAQmx on the calibration system. 
Driver support for calibrating the PXIe-4309 was first available in NI-DAQmx 17.1. For the list 
of devices supported by a specific release, refer to the 

NI-DAQmx Readme

, available on the 

version-specific download page or installation media.

You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx supports LabVIEW, 

LabWindows™/CVI™, C/C++, C#, and Visual Basic .NET. When you install NI-DAQmx, you 
only need to install support for the application software that you intend to use.

No other software is required to verify the operation of the TB-4309 (ST), TB-4309 (MT), or 
CAL-4309.

Summary of Contents for SC Express PXIe-4309

Page 1: ...tion 17 Specifications 17 Worldwide Support and Services 17 Software Calibrating the PXIe 4309 requires the installation of NI DAQmx on the calibration system Driver support for calibrating the PXIe 4...

Page 2: ...d reference information PXIe 4309 Specifications PXIe 4309 specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx NI DAQmx Help Informati...

Page 3: ...ut 0 1 V Output Impedance 50 DMM PXI 4071 or PXIe 4081 DC Voltage Input Range 10 V Accuracy 12 ppm of reading 0 5 ppm of range PXI Express Chassis PXIe 1062Q If this chassis is unavailable use another...

Page 4: ...re information refer to the Maintain Forced Air Cooling Note to Users document available at ni com manuals Allow adequate warm up time for all of the instruments and equipment according to the manufac...

Page 5: ...le for the calibration references Note The verification procedure requires performing a self calibration as a first step unless self verification is done immediately after an adjustment Refer to the S...

Page 6: ...Wait ten minutes for the thermal EMF to stabilize iv Connect the double banana plug shorting bar to the input of the CAL 4309 as shown in Figure 1 v Wait two minutes for the thermal EMF to stabilize F...

Page 7: ...p and clear the task g Compare the averaged values to the specified test limits in Table 2 1 TB 4309 ST Terminal Block 2 Aln Input Short C COPYRIGHT 2016 R FOR PATENTS NI COM PATENTS TB 4309 A1 B1 C1...

Page 8: ...bled 5 5 1 0 V 10 V 50 50 15 V 75 75 AI0 31 0 1 V Every Sample Disabled 4 4 1 0 V 10 V 15 V AI0 15 0 1 V None Enabled 2 2 1 0 V 10 V 15 V Verification of channels 0 through 31 provides coverages of ea...

Page 9: ...the PXIe 4309 to the calibrator output i Attach the CAL 4309 terminal block to the PXIe 4309 module ii Tighten the mounting screw at the top of the terminal block to secure the terminal block to the m...

Page 10: ...air nickel plated copper wires to AI0 through AI7 to allow the connection of the PXIe 4309 inputs to the calibrator one channel at a time iii Reinstall the top cover and tighten the top cover screws i...

Page 11: ...oltage channel task using the values in Table 5 c Start the task d Drop the first sample acquired for each channel e Average the remaining samples on a per channel basis f Stop and clear the task g Co...

Page 12: ...0 9999357 1 0000643 0 9999627 1 0000373 1 0 V 1 0000643 0 9999357 1 0000373 0 9999627 10 V 10 V 9 999445 10 000555 9 999715 10 000285 10 V 10 000555 9 999445 10 000285 9 999715 15 V 15 V 14 999019 15...

Page 13: ...ns required to adjust the PXIe 4309 Depending on the terminal block used for verification refer to step a or b for specifics about how to connect the calibrator output to the PXIe 4309 input 1 Measure...

Page 14: ...M input as shown in Figure 6 i Loosen the top cover screws and remove the top cover ii Attach the TB 4309 ST terminal block to the PXIe 4309 module iii Tighten the mounting screw at the top of the ter...

Page 15: ...pedance 10 G Auto Zero On ADC Calibration On DC Noise rejection High Order C COPYRIGHT 2016 R FOR PATENTS NI COM PATENTS TB 4309 A1 B1 C1 J12 AI28 AI28 AI12 AI12 AI20 AI20 AI4 AI4 AI29 AI29 AI13 AI13...

Page 16: ...mxAdjust4309Cal and enter the measured internal reference value 4 Call DAQmx Close External Calibration DAQmxCloseExtCal to close the session 5 Use the action cancel if there has been any error during...

Page 17: ...NI Factory Installation Services repairs extended warranty and other services Visit ni com register to register your NI product Product registration facilitates technical support and ensures that you...