OUTPut:PROTection:FOLDback
Foldback mode is used to disable the output when a transition is made between the operation
modes. The power supply will turn off the output after a specified delay if the power supply makes
transition into CV mode or into CC mode. This feature is particularly useful for protecting current
or voltage sensitive loads.
SCPI Command Syntax
OUTPut:PROT:FOLDback[:MODE] <CRD>
GEN Command Syntax
FLD <CRD>
Parameters
OFF|0, CC|1, CV|2 .
*RST Value
OFF
Examples
OUTPut:PROT:FOLDback[:MODE] CC
Query Syntax
OUTPut:PROT:FOLDback[:MODE]?
Returned Parameters
<CRD>
OUTPut:PROTection:DELay
Sets the delay time between the programming of an output change that produces a CV or CC
status condition. This command applies to UVP and Foldback functions.
SCPI Command Syntax
OUTPut:PROTection:DELay <NRf+>
GEN Command Syntax
FBD
Parameters
0.1 to 25.5|MIN|MAX (step 0.1s)
Unit
S (second)
*RST Value
0mS
Examples
OUTPut:PROTection:DELay 2E-1
Query Syntax
OUTPut:PROTection:DELay?
Returned Parameters
<NR3>
OUTPut:ILC:MODE
Selects the mode of operation of the Remote Inhibit protection. In OFF mode the power supply
ignores J3-4 (ILC) status.
SCPI Command Syntax
OUTPut:ILC:MODE <CRD>
GEN Command Syntax
RIE
Parameters
0|OFF 1|ON
*RST Value
OFF
Examples
OUTPut:ILC:MODE ON
Query Syntax
OUTPut:ILC:MODE?
Returned Parameters
ON/OFF