120
NATIONAL
INSTRUMENTS
ready for trigger and not in an idle state, send ABORt and then set INIT:CONT 0.
8.6.3 Wave Execution via Front Panel
1. Load stored data MENU -> ”
ROG
” -> ”
OAD
” ->
2
2. Set Trigger IN source ( BUS for Command or Front Panel and EXT via Rear Panel J3.2 pin)
MENU -> ”
TRiG
”
-> ”
TR. In
” -> ”
BS
”
3. Set Trigger Initialize Continue mode (If 1 then after program execution power supply will be
ready for the next trigger. If 0 then send INIT command MENU -> ”
TriG
” -> ”
ConT
” -> ”
EnA
” to
get to the next trigger.
4. Set INIT command. (Power supply ready for TRIGGER) MENU -> ”
TRiG
” -> ”
i niT
” -> ”
i niT
”
5. Exit from MENU and Press Current encoder to trigger.
8.7 Additional Examples
8.7.1 List Example
TRIG:SOUR BUS
Select trigger source external analog signal
0
2
4
6
8
10
0
1
2
3
4
5
Times [S]
Voltage [V]
VOLT:MODE LIST
Select sequence mode ”LIST”
LIST:VOLT 2,4,2,8,5,4
Set voltage values ”2,4,2,8,5,4” Volts
LIST:DWEL 0.5,0.5,1,1,1,1
Set dwell values ”0.5,0.5,1,1,1,1” seconds
LIST:COUN 1
Set list execution time ”1”
LIST:STEP AUTO
Set step execution mode ”AUTO”
INIT:CONT OFF
Trigger subsystem enabled for a single
trigger action
INIT
Trigger initializing
*TRG
Trigger command
8.7.2 Waveform Example
TRIG:SOUR BUS
Select trigger source via communication
interface or front panel
0
2
4
6
8
10
0
1
2
3
4
5
6
Times [S]
Voltage [V]
VOLT:MODE WAVE
Select sequence mode ”WAVE”
WAVE:VOLT 2,2,4,4,9,9,3,3
Set voltage values ”2,2,4,4,9,9,3,3” Volts
WAVE:TIME 0,1,0.5,0.5,0.5,0.5,1.5,1 Set time values ”0,1,0.5,0.5,0.5,0.5,1.5,1” seconds
WAVE:COUN 2
Set wave execution time ”2”
WAVE:STEP AUTO
Set step execution mode ”AUTO”
INIT:CONT ON
Trigger system is continuously enabled
INIT
Trigger initializing
*TRG
Trigger command
Summary of Contents for RMX-410 Series