If the installation window does not appear, navigate to the drive, double-click it, and
double-click
autorun.exe
.
3. Follow the instructions in the installation prompts.
Note
Windows users may see access and security messages during
installation. Accept the prompts to complete the installation.
4. When the installer completes, select
Restart
in the dialog box that prompts you to restart,
shut down, or restart later.
Installing the NI 5667
Caution
To prevent damage to the device caused by ESD or contamination, handle
the device using the edges or the metal bracket.
You must install NI-RFSA before installing the hardware.
Before you install the hardware, refer to the guidelines in the
Maintain Forced-Air Cooling
Note to Users
included with the module to ensure that the device can cool itself effectively.
To use the included cables, you must install the NI 5694 IF conditioning module immediately
to the left of the NI 5603 RF downconverter module, the NI 5622 IF digitizer module
immediately to the left of the NI 5694 module, the NI 5653 LO source module immediately to
the right of the NI 5603 RF downconverter module, and the NI 5693 RF preselector module
immediately to the right of the NI 5653 module.
1. Ensure the AC power source is connected to the chassis before installing the modules.
The AC power cord grounds the chassis and protects it from electrical damage while you
install the modules.
2. Power off the chassis.
3. Inspect the slot pins on the chassis backplane for any bends or damage prior to
installation. Do not install a module if the backplane is damaged.
4. Remove the black plastic connectors from all the captive screws on the module front
panel.
5. Identify a supported slot in the chassis. The following figure shows the symbols that
indicate the slot types.
NI PXIe-5667 (3.6 GHz) Getting Started Guide
|
© National Instruments
|
5
Summary of Contents for PXIe-5667