PXIe-5665 Calibration Procedure
Appendix B: Power Sensor Calibration Factor
Uncertainty
Table 56.
Combiner Paths
Combiner
Path
Start
Stop
Total
Loss
*
(dB, max.)
Isolation
†
(dB, min.)
VSWR
‡
(max.)
2
1
>700 MHz
10 GHz
8.0
20
1.75
2
18.0
2
1
>10 GHz
14 GHz
8.0
20
2.00
2
18.0
*
Total Loss
includes the splitter loss at either input or output ports. The unused input port is terminated
in 50
Ω
.
†
Isolation
is the loss between input ports. The output port is terminated in 50
Ω
.
‡
VSWR
includes any port. All other ports are terminated in 50
Ω
.
Table 57.
Maximum Power Sensor Calibration Factor Uncertainty
Frequency
Calibration Factor Uncertainty (%)
10 MHz
≤
1.15
20 MHz
≤
0.80
30 MHz
≤
0.65
40 MHz to 200 MHz
≤
0.55
300 MHz to 400 MHz
≤
0.60
500 MHz to 1 GHz
≤
0.50
2 GHz to 4 GHz
≤
0.60
5 GHz
≤
0.75
6 GHz
≤
0.80
7 GHz
≤
0.85
8 GHz
≤
0.90
9 GHz
≤
0.95
10 GHz
≤
1.00