8.
Repeat steps 5 through 7 until the
OCXO calibration complete
output of the niRFSG
5654 OCXO Cal Adjust VI returns a value of TRUE.
9.
Call the niRFSG Close External Calibration VI to close the session. Set the
write
calibration to hardware?
parameter to TRUE to store the results to the EEPROM on the
PXIe-5654.
Adjusting RF OUT Power
This procedure adjusts the PXIe-5654 RF OUT power using a power meter. This adjustment
yields a more accurate output power for the PXIe-5654 RF OUT front panel connectors.
1.
Connect the power sensor to the PXIe-5654 RF OUT front panel connector as shown in
the following figure.
Figure 12. RF OUT Power Adjustment Equipment Setup
250 kHz–20 GHz Signal Generator
ACCESS
ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50
Ω
2
1
43
1. PXIe-5654 RF Analog Signal Generator
2. Power Meter
3. Power Sensor
2.
Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note
The default password for password-protected operations is
NI
.
3.
Call the niRFSG 5654 Power Cal Initialize VI.
4.
Call the niRFSG 5654 Power Cal Configure VI.