NI 5622 Calibration Procedure
30
ni.com
Timing Accuracy
Complete the following steps to verify the timing accuracy of the NI 5622.
1.
Connect the signal generator to channel 0 of the digitizer.
2.
Open a session and obtain a session handle using the niScope
Initialize VI.
3.
Configure the input impedance and the maximum input frequency
using the niScope Configure Chan Characteristics VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the following
parameters:
resourceName
: The device name assigned by
MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the following
parameters:
vi
: The instrument handle from
niScope_init
inputImpedance
:
NISCOPE_VAL_50_OHM
maxInputFrequency
:
250,000,000