4.
Enable the calibrator output if not already enabled.
Note
To configure the calibrator to output 0V, enable ground selection.
Otherwise, disable ground selection to generate a nonzero voltage.
5.
Wait 1 second for settling.
6.
Call the niHsai Group A External Cal API v1 Host.lvclass:DC Reference Cal Adjust VI
with the following settings:
•
actual voltage generated (V)
: The DC voltage present on the specified channel of
the PXIe-5172
7.
Repeat steps 3 through 6 until the
DC cal complete
indicator from the niHsai Group A
External Cal API v1 Host.lvclass:DC Reference Cal Adjust VI returns TRUE.
8.
Disable the calibrator output.
Return to the main
task.
Adjusting Timebase
Follow this procedure to adjust the internal timebase reference of the PXIe-5172.
1.
Connect the calibrator test head to the specified channel of the PXIe-5172 using an SMB
(plug) to BNC (f) adapter.
2.
Call the niHsai Group A External Cal API v1 Host.lvclass:Timebase Cal Initialize VI
with the following settings:
•
channel
: The specified channel
3.
Call the niHsai Group A External Cal API v1 Host.lvclass:Timebase Cal Configure VI to
obtain the frequency to generate and configure the calibrator to output a 1.25 V
pk-pk
sine
wave at the specified frequency into 50 Ω.
4.
Enable the calibrator output if not already enabled.
5.
Wait 1 second for settling.
6.
Call the niHsai Group A External Cal API v1 Host.lvclass:Timebase Cal Adjust VI with
the following settings:
•
actual frequency generated (Hz)
: The frequency of the sine wave present on the
specified channel of the PXIe-5172
7.
Repeat steps 3 through 6 until the
Timebase Cal Complete
indicator from the Timebase
Cal Adjust VI returns TRUE.
8.
Disable the calibrator output.
Return to the main
task.
Reverification
section to determine the as-left status of the PXIe-5172.
Note
If any test fails reverification after performing an adjustment, verify that you
have met the test conditions before returning your PXIe-5172 to NI. Refer to the
36
|
ni.com
|
PXIe-5172 Calibration Procedure