17. Connect splitter output 2 of the power sensor assembly from the
section to channel 1 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
18. Disable the signal generator output.
19. Connect splitter output 2 of the power sensor assembly from the
section to channel 2 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
20. Disable the signal generator output.
21. Connect splitter output 2 of the power sensor assembly from the
section to channel 3 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
22. Disable the signal generator output.
Note
If you are verifying the PXIe-5172 (8CH), proceed to the next step. If
you are verifying the PXIe-5172 (4CH), 1 MΩ Bandwidth Verification is
complete.
23. Connect splitter output 2 of the power sensor assembly from the
section to channel 4 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
24. Disable the signal generator output.
25. Connect splitter output 2 of the power sensor assembly from the
section to channel 5 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
26. Disable the signal generator output.
27. Connect splitter output 2 of the power sensor assembly from the
section to channel 6 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
28. Disable the signal generator output.
29. Connect splitter output 2 of the power sensor assembly from the
section to channel 7 of the PXIe-5172. Repeat steps 3 through 14 for
each configuration in the
1 MΩ Bandwidth Verification Table
30. Disable the signal generator output.
Verifying Timebase Accuracy
Follow this procedure to verify the frequency accuracy of the PXIe-5172 onboard timebase
using an oscilloscope calibrator.
Table 7. Timebase Accuracy Verification
As-Found Limits
As-Left Limits
Measurement Uncertainty
5
±25.0 PPM (±2,225 Hz)
±10.0 PPM (±890 Hz)
±0.6 PPM (±54 Hz)
5
Measurement uncertainty based on Fluke 9500B with Fluke 9530 test head specifications that
apply at T
cal
±5 °C, where Factory T
cal
= 23 °C. Uncertainty of the 9500B includes long-term
30
|
ni.com
|
PXIe-5172 Calibration Procedure