PXIe-4468 Calibration Procedure
4
ni.com/services
Calibration Overview
Recommended Calibration Interval
2 years
Password
NI
Note
This is the default password for all password-protected operations.
This password is site-specific.
Task
Estimated Test
Time
Operator
Connections
Test Points
Setup
10 minutes
—
—
Warm Up
15 minutes
—
—
Verify, Adjust, and
Reverify
19 minutes
36
344
Verify Only
8 minutes
17
171
Adjust Only
3 minutes
2
2
Note
Estimated test times assume the user is conducting a manual
calibration. For most procedures, automating the calibration
significantly reduces test times.
Environmental Conditions
Verification
Adjustment
Ambient temperature
23
°C ±
5
°C
23
°C ±
5
°C
Internal DUT temperature range
Tcal
±
5
°C
Tcal
±
5
°C
Relative humidity
Between 20% and 80%, noncondensing