Table 5.
NI PXIe-4353 Offset with Autozero Enabled Verification Test Limits
Timing Mode
Test Point
1-Year Limits (V)
Location
Value (V)
Lower Limit
Upper Limit
1 (High Resolution)
Mid
0.000000
–0.000002
0.000002
2
–0.000002
0.000002
3
–0.000002
0.000002
4
–0.000003
0.000003
5
–0.000005
0.000005
6
–0.0000065
0.0000065
7 (High Speed)
–0.0000075
0.0000075
Table 6.
NI PXIe-4353 Offset with Autozero Disabled Verification Test Limits
Timing Mode
Test Point
1-Year Limits (V)
Location
Value (V)
Lower Limit
Upper Limit
1 (High Resolution)
Mid
0.000000
–0.0000095
0.0000095
2
–0.0000095
0.0000095
3
–0.0000095
0.0000095
4
–0.0000105
0.0000105
5
–0.000013
0.000013
6
–0.0000145
0.0000145
7 (High Speed)
–0.0000155
0.0000155