© National Instruments
|
2-7
Self-Calibration
The PXIe-4340 can adjust its ratiometric gain and offset to cancel out the drift of the analog
circuitry. In addition, self-calibration nulls the DC component of the excitation source and
adjusts its amplitude against an internal reference. The PXIe-4340 does not support external
adjustment.
Module Pinout
Table 2-1 shows the pinout of the front connector of the PXIe-4340. Refer to the
I/O Connector Signal Descriptions
section for definitions of each signal. Refer to the
NI PXIe-4340 and TB-4340 User Guide and Terminal Block Specifications
for signal locations