fault, interruption of the current can be done by programmatically disconnecting EXT PWR
from the appropriate position by stopping the tasks that are using EXT PWR+. The following
three steps summarize the steps that should be taken to reset a tripped PTC:
1.
Remove the short-circuit fault from the affected channel.
2.
Programmatically disconnect EXT PWR from the position of which the faulted channel
was a part.
3.
After 10 seconds, programmatically reconnect EXT PWR.
Position Resources
This section describes the position resources on both the front and rear of the RM-4339. Figure 5
shows the locations of these resources.
RJ50
Refer to Table 5 for signal names and descriptions, and the correlation to the pin numbers of the
RJ50 (10p10c) modular plug. Refer to Figure 8 for default switch positions.