NI PXIe-4339 Calibration Procedure
|
© National Instruments
|
29
17. Call the 4339 instance of the DAQmx Adjust SC Express Calibration function with the
following parameters:
calhandle in
: calhandle output from DAQmx Initialize External Calibration
action
: commit
0.2
-0.2
2.5
0.45
0
-0.45
Table 24.
Ratio Mode (±4 V/V Range)
Range (V/V)
Excitation Setting (V)
Calibrator Output (V)
Max
Min
4
-4
0.625
2.25
0
-2.25
4
-4
1
3.6
0
-3.6
4
-4
1.5
5.4
0
-5.4
4
-4
2
7.2
0
-7.2
4
-4
2.5
9
0
-9
Table 23.
Ratio Mode (±0.2 V/V Range) (Continued)
Range (V/V)
Excitation Setting (V)
Calibrator Output (V)
Max
Min