National Instruments PXIe-4302 Calibration Procedure Download Page 14

© 2015–2021 National Instruments Corporation. All rights reserved. 

377005B-01

Sep21

Information is subject to change without notice. Refer to the 

NI Trademarks and Logo Guidelines 

at 

ni.com/trademarks

 for more 

information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective 
companies. For patents covering NI products/technology, refer to the appropriate location: 

Help»Patents

 in your software, the 

patents.txt

 

file on your media, or the 

National Instruments Patents Notice

 at 

ni.com/patents

. You can find information about end-user license 

agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the 

Export Compliance Information 

at 

ni.com/legal/export-compliance

 for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and 

other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION 
CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual 
was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 
252.227-7014, and DFAR 252.227-7015.

Specifications

Refer to the 

NI PXIe-4302/4303 Specifications 

document for detailed PXIe-4302/4303 

specification information.

Refer to the 

NI PXIe-4302/4303 and TB-4302/4302C User Guide and Terminal Block 

Specifications

 document for detailed TB-4302C specification information.

Revision History

NI Services

Visit 

ni.com/support

 to find support resources including documentation, downloads, and 

troubleshooting and application development self-help such as tutorials and examples.

Visit 

ni.com/services

 to learn about NI service offerings such as calibration options, repair, 

and replacement.

Visit 

ni.com/register

 to register your NI product. Product registration facilitates technical 

support and ensures that you receive important information updates from NI. 

NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 
78759-3504, USA. 

Part Number

Edition Date

Section

Changes

377005B-01

Sept 2021

Accuracy Verification

Updated limits

Adjustment

Updated the calibration 
function parameter

Summary of Contents for PXIe-4302

Page 1: ...date 9 Reverification 10 TB 4302C Verification 10 Test Equipment 10 Accuracy Verification 11 Specifications 14 Revision History 14 NI Services 14 Software Calibrating the PXIe 4302 4303 requires you t...

Page 2: ...ting the PXIe 4302 4303 Test Equipment Table 1 lists the equipment recommended for the performance verification and adjustment procedures of the PXIe 4302 4303 If the recommended equipment is not avai...

Page 3: ...SMU PXIe 4139 Noise 0 1 Hz to 10 Hz peak to peak is 60 V or better at 10 V Noise 0 1 Hz to 10 Hz peak to peak is 2 V or better at 100 mV Table 1 Recommended Equipment for PXIe 4302 4303 Verification a...

Page 4: ...th 1 or better tolerance to build a voltage divider to bias the PXIe 4139 output and set the common mode input of the PXIe 4302 4303 to zero volts Connect one resistor between AI and AIGND and the oth...

Page 5: ...n Guide and Terminal Block Specifications for information about how to install the software and hardware and how to configure the device in Measurement Automation Explorer MAX Note When a device is co...

Page 6: ...Voltage output Sense Remote Range 600 mV range for test points less than 100 mV 60 V range for all other test points Current Limit 20 mA Current Limit Range 200 mA Table 4 PXI 4071 Voltage Measurement...

Page 7: ...onfiguration Value Channel Name Dev1 ai0 32 Task AI voltage Sample Mode Finite samples Sample Clock Rate 5000 Samples per Channel 5000 Maximum Value Appropriate maximum range value from Table 6 Minimu...

Page 8: ...with the following parameters Device In Dev1 Password NI1 4 Call the 4302 4303 instance of the DAQmx Setup SC Express Calibration function with the following parameters calhandle in calhandle output f...

Page 9: ...eps 5 through 8 for the remaining Test Point values from Table 7 for the corresponding range that was configured in step 4 10 Repeat steps 4 through 9 for the remaining ranges from Table 7 11 Call the...

Page 10: ...returning the device to NI TB 4302C Verification This section provides information for verifying the performance of the TB 4302C Test Equipment Table 8 lists the equipment recommended for verifying t...

Page 11: ...VSUP VSUP A9 AI0 VSUP VSUP AI19 VSUP AI18 VSUP AI20 AI21 AI23 VSUP AI22 VSUP VSUP VSUP AI16 AI17 VSUP VSUP AI28 VSUP AI27 VSUP AI29 AI30 AI31 VSUP AI24 VSUP VSUP VSUP A25 AI26 VSUP VSUP S N 158141A 0...

Page 12: ...annel to Shunt Reference Designator Correlation Channel Shunt Reference Designator CH0 R10 CH1 R11 CH2 R12 CH3 R13 CH4 R14 CH5 R15 CH6 R16 CH7 R17 CH8 R21 CH9 R20 CH10 R19 CH11 R18 CH12 R25 CH13 R24 C...

Page 13: ...1 6 Compare the results to the Lower Limit and Upper Limit values in Table 11 If the results are between these values the device passes the test 7 Repeat steps 3 through 6 for all other 5 shunt resist...

Page 14: ...Customers The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014...