Page 1: ...PXI Express TM PXIe 1095 User Manual PXIe 1095 User Manual March 2018 377169B 01...

Page 2: ...port phone numbers email addresses and current events National Instruments Corporate Headquarters 11500 North Mopac Expressway Austin Texas 78759 3504 USA Tel 512 683 0100 For further support informat...

Page 3: ...ESS FOR A PARTICULAR PURPOSE TITLE OR NON INFRINGEMENT AND ANY WARRANTIES THAT MAY ARISE FROM USAGE OF TRADE OR COURSE OF DEALING NI DOES NOT WARRANT GUARANTEE OR MAKE ANY REPRESENTATIONS REGARDING TH...

Page 4: ...es independent from National Instruments and have no agency partnership or joint venture relationship with National Instruments Patents For patents covering National Instruments products technology re...

Page 5: ...ntroller Slot 1 6 Hybrid Peripheral Slots 1 7 PXI Express Peripheral Slots 1 7 System Timing Slot 1 8 PXI Local Bus 1 9 PXI Trigger Bus 1 9 System Reference Clock 1 10 Chapter 2 Installation and Confi...

Page 6: ...X 2 16 Using System Configuration and Initialization Files 2 17 Chapter 3 Maintenance Service Interval 3 1 Preparation 3 1 Cleaning 3 1 Interior Cleaning 3 1 Exterior Cleaning 3 2 Replacing the Power...

Page 7: ...nd helpful as you read this manual IEEE 1101 1 1991 IEEE Standard for Mechanical Core Specifications for Microcomputers Using IEC 603 2 Connectors IEEE 1101 10 IEEE Standard for Additional Mechanical...

Page 8: ...he inner chassis for any possible damage debris or detached components If damage appears to have been caused during shipment file a claim with the carrier Retain the packing material for possible insp...

Page 9: ...The key features of the PXIe 1095 chassis include the following High Performance for Instrumentation Requirements Up to 8 GB s single direction per PXI Express slot dedicated bandwidth x8 Gen 3 PCI E...

Page 10: ...tchless CLK10 routing Optional Features Timing and synchronization upgrade Front and rear rack mount kits Replacement power supply EMC filler panels Slot blockers for improved cooling performance Fact...

Page 11: ...heral Slots 11x 5 PXI Express Hybrid Peripheral Slots 5x 6 PXI Express System Timing Slot 7 PXI Express System Controller Slot 8 Front Panel LEDs 9 Power Inhibit Switch 10 DIP Switch 11 Chassis Carry...

Page 12: ...vailable from National Instruments You easily can install replacement power supplies without powering off the system Replacement Fan Kit A fan kit is available from National Instruments includes both...

Page 13: ...d to PCI Express switch 1 with Gen 3 x8 PCI Express links and are downstream of system slot link 1 The PCI Express to PCI bridge is connected to PCI Express switch 1 and provides a 32 bit 33 MHz PCI b...

Page 14: ...PXIExpressfunctionalityand32 bitPXIfunctionalityexcept for PXI Local Bus The hybrid peripheral slot only connects to PXI Local Bus 6 left and right PXI Express Peripheral Slots There are eleven 11 PXI...

Page 15: ...or for each PXI Express peripheral or hybrid peripheral slot as well as routed back to the XP3 connector of the system timing slot as shown in Figure 1 4 The PXIe_DSTAR pairs can be used for high spee...

Page 16: ...control carefully timed sequences of operations performed on other modules in the system Modules can pass triggers to one another allowing precisely timed responses to asynchronous external events the...

Page 17: ...LVPECL termination for the buffer to drive PXIe_SYNC100 so that when there is no peripheral or a peripheral that does not connect to PXIe_SYNC100 there is no SYNC100 signal being driven on the pair t...

Page 18: ...connecting to AC power The facility installation shall provide a means for connection to protective earth and Qualified personnel shall install a protective earthing conductor from the chassis protect...

Page 19: ...of the chassis The module intake vents are on the rear of the chassis There are also intake and exhaust vents located along the sides of the chassis The vent locations are shown in Figure 2 2 PXIe 10...

Page 20: ...PXIe 1095 User Manual Figure 2 1 PXIe 1095 Chassis Minimum Cooling Clearances 101 60 mm 4 00 in 44 45 mm 1 75 in 44 45 mm 1 75 in PS TEMP FANS PXIe 1095 USE ONLY COMPATIBLE RACK MOUNT KITS COOLING CLE...

Page 21: ...cated on the left side of the chassis 1 PXI Module Air Intake 3x 2 Power Supply Intake 2x 3 PXI Module Air Exhaust Vent 4 Power Supply Air Exhaust Vent 2x 5 Timing and Synchronization Upgrade Air Exha...

Page 22: ...The PXIe 1095 chassis supports multiple fan operating modes Refer to the Fan Mode section for more information Power Supply Filler Panel To maintain proper chassis cooling performance a power supply f...

Page 23: ...from National Instruments Refer to the instructions supplied with the rack mount kits to install your PXIe 1095 chassis in an instrument rack Note You may want to remove the feet or carrying handle f...

Page 24: ...a grounding lug The wire must have green insulation with a yellow stripe or must be noninsulated bare 2 Attach the opposite end of the wire to permanent earth ground using toothed washers or a toothed...

Page 25: ...nd bottom Slide the system controller to the rear of the chassis making sure that the injector ejector handle is pushed down as shown in Figure 2 4 Figure 2 4 Installing a PXI Express System Controlle...

Page 26: ...ssis before installing the module The AC power cord grounds the chassis and protects it from electrical damage while you install the module 2 Ensure that the chassis is powered off 3 Install a module...

Page 27: ...sis power supply or supplies are active and operating normally Blinking red In a redundant power setup one power supply has failed Steady red The chassis power supply or supplies have failed Temperatu...

Page 28: ...s switch is in the off right position Auto mode is selected When this switch is in the on left position High mode is selected DIP switch 2 second from the bottom controls the chassis Inhibit Mode When...

Page 29: ...or floating it will cause the chassis to be powered on Driving the Remote Inhibit signal low or shorting it to ground will cause main power to be inhibited Note The Timing and Synchronization upgrade...

Page 30: ...h on the backplane Refer to the DIP Switches section for more information about the DIP switch Refer to Figure 1 1 Front View of the PXIe 1095 Chassis for the location of this switch Note The DIP swit...

Page 31: ...2 14 ni com Chapter 2 Installation and Configuration Number and type of power supplies Slot details Chassis self test Firmware Update Figure 2 8 Chassis Settings in MAX...

Page 32: ...ic reservations and routing are used whenever possible If static reservations are required static reservation of trigger lines can be implemented by the user in MAX through the Triggers tab Reserved t...

Page 33: ...e triggers on that line in the described direction or select Dynamic for the default behavior with no manual routing 4 Click the Save button Fan Configuration in MAX You can configure fan behavior usi...

Page 34: ...the system controller to create a single system initialization file called pxisys ini PXI System Initialization The system controller manufacturer either provides a pxisys ini file for the particular...

Page 35: ...ocedures in this chapter Caution Many components within the chassis are susceptible to static discharge damage Service the chassis only in a static free environment Observe standard handling precautio...

Page 36: ...moisture to dampen the cloth Do not wash the front or rear panel connectors or switches Cover these components while cleaning the chassis Do not use harsh chemical cleaning agents they may damage the...

Page 37: ...y to the chassis Refer to Figure 3 1 Removing PXIe 1095 Power Supply for the screw locations Using a Phillips screwdriver remove the screws Pull on the rear handle of the power supply to remove it fro...

Page 38: ...Specifications tighten screws to 11 5 in lb 1 3 N m of torque Connecting Safety Ground Refer to the Connecting the Safety Ground section of Chapter 2 Installation and Configuration Connecting to Powe...

Page 39: ...3 With internal fan harness still connected carefully pull and rotate fan assembly from rear cavity of chassis Use caution when removing the fan assembly to avoid damaging the fan wire harness 4 Disc...

Page 40: ...ce Figure 3 2 Replacing Rear Fan Module Figure 3 3 Internal Fan Harness 1 Fan Harness Plug 2 PXIe 1095 Chassis 3 PXI Module Fan Assembly 4 Rear Chassis Feet 2x 5 Mounting Screws 8x 1 Fan Receptacle 2...

Page 41: ...sconnect the fan from the chassis receptacle Use caution when removing the fan assembly to avoid damaging the internal wire harness 5 Pull side fan assembly straight from chassis and remove Complete t...

Page 42: ...Figure 3 4 Replacing Side Fan Assembly 1 Chassis Side Fan Cavity 2 Side Fan Assembly 3 Side Fan Cover 4 Side Fan Cover Mounting Screws 4x 1 4 2 3...

Page 43: ...nnector Pinout for the System Controller slot Table A 6 shows the TP2 Connector Pinout for the System Timing slot Table A 7 shows the XP3 Connector Pinout for the System Timing slot Table A 8 shows th...

Page 44: ...ETp3 2PETn3 GND 2PERp3 2PERn3 GND 2PERp2 2PERn2 GND 3 2PETp4 2PETn4 GND 2PERp4 2PERn4 GND 2PETp5 2PETn5 GND 4 2PETp6 2PETn6 GND 2PERp6 2PERn6 GND 2PERp5 2PERn5 GND 5 2PETp7 2PETn7 GND 2PERp7 2PERn7 GN...

Page 45: ...1PERn1 GND 7 1PETp3 1PETn3 GND 1PERp3 1PERn3 GND 1PETp4 1PETn4 GND 8 1PETp5 1PETn5 GND 1PERp5 1PERn5 GND 1PERp4 1PERn4 GND 9 1PETp6 1PETn6 GND 1PERp6 1PERn6 GND 1PETp7 1PETn7 GND 10 2PETp0 2PETn0 GND...

Page 46: ...C NC GND 3 PXIe_DSTARB4 PXIe_DSTARB4 GND NC NC GND NC NC GND 4 PXIe_DSTARA4 PXIe_DSTARA4 GND NC NC GND NC NC GND 5 PXIe_DSTARC5 PXIe_DSTARC5 GND PXI_STAR14 PXI_STAR15 GND NC NC GND 6 PXIe_DSTARB5 PXIe...

Page 47: ...e_DSTARB1 PXIe_DSTARB1 GND PXI_STAR0 PXI_STAR1 GND NC NC GND 5 PXIe_DSTARA1 PXIe_DSTARA1 GND PXI_STAR2 PXI_STAR3 GND NC NC GND 6 PXIe_DSTARC2 PXIe_DSTARC2 GND PXI_STAR4 PXI_STAR5 GND NC NC GND 7 PXIe_...

Page 48: ...PXIe_DSTARA PXIe_DSTARA GND 3 SMBDAT SMBCLK GND RSV RSV GND RSV RSV GND 4 MPWRGD PERST GND RSV RSV GND 1RefClk 1RefClk GND 5 1PETp0 1PETn0 GND 1PERp0 1PERn0 GND 1PETp1 1PETn1 GND 6 1PETp2 1PETn2 GND 1...

Page 49: ...tor Pinout for the Hybrid Slot Pin Z A B C D E F 25 GND 5V REQ64 ENUM 3 3V 5V GND 24 GND AD 1 5V V I O AD 0 ACK64 GND 23 GND 3 3V AD 4 AD 3 5V AD 2 GND 22 GND AD 7 GND 3 3V AD 6 AD 5 GND 21 GND 3 3V A...

Page 50: ...1A5 BRSVP1B5 RST GND GNT GND 4 GND IPMB_PWR HEALTHY V I O INTP INTS GND 3 GND INTA INTB INTC 5V INTD GND 2 GND TCK 5V TMS TDO TDI GND 1 GND 5V 12V TRST 12V 5V GND Table A 9 P1 Connector Pinout for the...

Page 51: ...n2 GND 1PERp2 1PERn2 GND 1PERp1 1PERn1 GND 7 1PETp3 1PETn3 GND 1PERp3 1PERn3 GND 1PETp4 1PETn4 GND 8 1PETp5 1PETn5 GND 1PERp5 1PERn5 GND 1PERp4 1PERn4 GND 9 1PETp6 1PETn6 GND 1PERp6 1PERn6 GND 1PETp7...

Page 52: ...you identify your systems accuracy and reliability requirements and provides warranty sparing and calibration services to help you maintain accuracy and minimize downtime over the life of your system...

Page 53: ...tions Engineers make sure every question submitted online receives an answer Software Support Service Membership The Standard Service Program SSP is a renewable one year subscription included with alm...

Page 54: ...10 9 micro 10 6 m milli 10 3 k kilo 103 M mega 106 G giga 109 T tera 1012 Symbols Degrees Equal or greater than Equal or less than Percent A A Amperes AC Alternating current ANSI American National St...

Page 55: ...y standard mechanical components and high performance connector technologies to provide an optimized system intended for rugged applications It is electrically compatible with the PCI Specification wh...

Page 56: ...cceleration equal to 9 8 m s2 GPIB General Purpose Interface Bus IEEE 488 gRMS A measure of random vibration The root mean square of acceleration levels in a random vibration test profile H hr Hours H...

Page 57: ...age that a DC voltage output will change as a result of a specified change in input AC voltage step change from 90 to 132 VAC or 180 to 264 VAC load regulation The maximum steady state percentage that...

Page 58: ...cker An assembly installed into an empty slot to improve the airflow in adjacent slots SMA SubMiniature version A connector a commonly used coaxial connector standby The backplane is unpowered off but...

Page 59: ...s in a measurement or control system The 10 MHz REF IN and OUT BNC connectors on the rear of the chassis can be used to synchronize multiple chassis to one reference clock The PXI backplane specificat...

Page 60: ...el installation 2 5 slot blocker installation 2 5 E EMC filler panel kit 1 5 F fan module replacing 1 5 filler panel installation 2 5 G ground connecting 2 6 H hybrid peripheral slots description 1 7...

Page 61: ...nce clock 1 10 system timing slot 1 8 trigger bus 1 9 PXIe 1095 chassis installation See installation configuration and operation key features 1 2 maintenance See maintenance of PXIe 1095 chassis opti...