GETTING STARTED GUIDE

NI PXI-5671

2.7 GHz RF Vector Signal Generator

Note

 

Before you begin, install and configure your chassis and controller.

This document explains how to install, configure, and test the NI PXI-5671 (NI 5671). The
NI 5671 is a 2.7 GHz RF vector signal generator (VSG) with digital upconversion. The
NI 5671 ships with the NI-RFSG instrument driver, which you use to program the device.

The NI 5671 comprises the following devices:

NI PXI-5610 (NI 5610) RF upconverter module

NI PXI-5441 (NI 5441) arbitrary waveform generator (AWG) module

To access NI 5671 documentation, navigate to 

Start

»

All Programs

»

National Instruments

»

NI-RFSG

»

Documentation

.

The specifications document for your device is installed with the driver software. Refer to 

ni.com/manuals

 for the most recent specifications for your device.

Caution

 

The protection provided by this product may be impaired if it is used in a

manner not described in this document.

Hot Surface

 

If the NI 5671 has been in use, it may exceed safe handling

temperatures and cause burns. Allow the NI 5671 to cool before removing it from
the chassis.

Contents

Electromagnetic Compatibility Guidelines...............................................................................2
Verifying the System Requirements..........................................................................................2
Unpacking the Kit..................................................................................................................... 2

Verifying the Kit Contents................................................................................................ 4

Preparing the Environment....................................................................................................... 5
Installing the Software.............................................................................................................. 6
Installing the NI 5671............................................................................................................... 6

Interconnecting the NI 5671 Modules.............................................................................. 8

Configuring the NI 5671 in MAX...........................................................................................15

Summary of Contents for PXI-5671

Page 1: ...rt All Programs National Instruments NI RFSG Documentation The specifications document for your device is installed with the driver software Refer to ni com manuals for the most recent specifications...

Page 2: ...al device or test object or if the product is used in residential or commercial areas To minimize interference with radio and television reception and prevent unacceptable performance degradation inst...

Page 3: ...s or any other sign of damage Caution Never touch the exposed pins of connectors Notify NI if the device appears damaged in any way Do not install a damaged device 3 Unpack any other items and documen...

Page 4: ...wo Semi Flexible SMA m to SMB Cables Part Number 190846B 01 4 Two Cable Wrenches Part Number 746016 01 5 Screwdriver Part Number 772006 01 6 Driver Software DVD 7 NI PXI 5671 Getting Started Guide thi...

Page 5: ...ets the following specifications Operating ambient temperature NI PXI 5610 upconverter 0 C to 55 C Tested in accordance with IEC 60068 2 1 and IEC 60068 2 2 Meets MIL PRF 28800F Class 3 low temperatur...

Page 6: ...NI RFSG before installing the hardware Before you install the hardware refer to the guidelines in the Maintain Forced Air Cooling Note to Users included with the module to ensure that the chassis can...

Page 7: ...and bottom of the chassis Slide the device into the slot until it is fully inserted Figure 3 Module Installation 2 3 NI PXIe 1075 1 1 Chassis 2 Hardware Module 3 Ejector Handle in Down Unlatched Posit...

Page 8: ...ATA CONTROL 2 1 1 Flexible SMA to SMB Cable 2 Flexible SMA to SMB Cable 1 Using a flexible SMA to SMB cable connect the INPUT connector on the NI 5610 front panel to the CH 0 connector on the NI 5441...

Page 9: ...type material and manufacturer 5 Power on your PXI chassis and controller system 6 Verify that the ACCESS LED is illuminated on all modules If an LED is not illuminated refer to the Troubleshooting s...

Page 10: ...two multicolor LEDs Figure 5 NI 5610 RF Upconverter Module Front Panel Table 1 Device Front Panel Icon Definition Refer to the user documentation for required maintenance measures to ensure user safet...

Page 11: ...her The output signal at these connectors is always on and cannot be disabled 10 MHz OUT REF IN Routes an external frequency reference signal that can be propagated to the PXI backplane The NI 5610 ca...

Page 12: ...he device GREEN The module is ready to be programmed by NI RFSG ACTIVE Indicates the status of the module PLLs OFF The module is in an uninitialized state there is no error AMBER The module PLLs are a...

Page 13: ...Module Front Panel PFI 0 PFI 1 NI PXI 5441 ACCESS ACTIVE CLK IN CH 0 DIGITAL DATA CONTROL Table 4 Device Front Panel Icon Definition Refer to the user documentation for required maintenance measures t...

Page 14: ...trigger outputs trigger inputs and a clock input Table 6 NI 5441 AWG Module Front Panel LEDs LED Indications ACCESS Indicates the basic hardware status of the module OFF The module is not yet function...

Page 15: ...the device is correctly installed and restart 4 Record the device identifier MAX assigns to the hardware Use this identifier when programming the NI 5671 5 Associate the hardware modules that compris...

Page 16: ...es of the NI 5671 in MAX NI RFSG operates the modules as a single instrument LabWindows CVI Available at Program Files IVI Foundation IVI Drivers niRFSG Microsoft Visual C C Add all required include a...

Page 17: ...shooting section of this document if an ACTIVE LED does not turn on or if the NI RFSG SFP generates an error During signal generation the ACTIVE LED on each of the NI 5671 hardware modules illuminates...

Page 18: ...anel 11 In the VI front panel power level dBm control enter 0 In the frequency Hz control enter 100M 100 MHz 12 In the VI front panel resource name control enter the NI 5610 upconverter device name th...

Page 19: ...of the status element to an input of the Or function 9 Right click the unused input of the Or function and select Create Control to create a Boolean control 10 In the VI front panel right click the Bo...

Page 20: ...erter module name specified in MAX in the resource name control 4 Click the Run button on the toolbar to initiate sine wave generation 5 Click the VI front panel STOP button to stop sine wave generati...

Page 21: ...install a damaged device 4 Reinstall the module in a different chassis slot 5 Power on the chassis 6 Verify that the device appears in MAX 7 Reset the device in MAX and perform a self test What Should...

Page 22: ...y cooled 3 Reset the device in MAX The thermal shutdown error continues to be reported until you successfully reset the device What Should I Do if I Experience Performance Issues Using an MXI Connecti...