Storage Environment
Ambient temperature range
-20 °C to 70 °C (Tested in accordance
with IEC 60068-2-1 and IEC 60068-2-2.)
Relative humidity range
5% to 95%, noncondensing (Tested in
accordance with IEC 60068-2-56.)
Shock and Vibration
Operational shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC 60068-2-27. Test profile
developed in accordance with
MIL-PRF-28800F.)
Random vibration
Operating
5 Hz to 500 Hz, 0.31 g
rms
(Tested in
accordance with IEC 60068-2-64.)
Nonoperating
5 Hz to 500 Hz, 2.46 g
rms
(Tested in
accordance with IEC 60068-2-64. Test profile
exceeds the requirements of MIL-PRF-28800F,
Class 3.)
Compliance and Certifications
Safety
This product is designed to meet the requirements of the following electrical equipment safety
standards for measurement, control, and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the
section.
Electromagnetic Compatibility
This product meets the requirements of the following EMC standards for electrical equipment
for measurement, control, and laboratory use:
•
EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity
•
EN 55011 (CISPR 11): Group 1, Class A emissions
•
AS/NZS CISPR 11: Group 1, Class A emissions
NI PXI-2568 Specifications
|
© National Instruments
|