background image

Artisan Technology Group 

is your source for quality 

new and certified-used/pre-owned equipment

• 

FAST SHIPPING AND 

DELIVERY

• 

TENS OF THOUSANDS OF 

IN-STOCK ITEMS

•  EQUIPMENT DEMOS
• 

HUNDREDS OF 

MANUFACTURERS 

SUPPORTED

•  LEASING/MONTHLY

 

RENTALS

•  ITAR CERTIFIED

           

SECURE ASSET 

SOLUTIONS

SERVICE CENTER REPAIRS

Experienced engineers and technicians on staff 

at our full-service, in-house repair center

WE BUY USED EQUIPMENT

Sell your excess, underutilized, and idle used equipment 

We also offer credit for buy-backs and trade-ins

www.artisantg.com/WeBuyEquipment

                        

REMOTE INSPECTION

Remotely inspect equipment before purchasing with 

our interactive website at

 

www.instraview.com

LOOKING FOR MORE INFORMATION? 

Visit us on the web at

 

www.artisantg.com

    

for more 

information on price quotations, drivers, technical 

specifications, manuals, and documentation

Contact us:

 

 

(888) 88-SOURCE  |  [email protected]  |  www.artisantg.com

SM

View

Instra

Summary of Contents for PXI-1010

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...Manual Combination Chassis for PXI CompactPCI and SCXI Modules NI PXI 1010 Chassis User Manual May 2004 Edition Part Number 371249A 01 Artisan Technology Group Quality Instrumentation Guaranteed 888...

Page 3: ...0 51190000 Israel 972 0 3 6393737 Italy 39 02 413091 Japan 81 3 5472 2970 Korea 82 02 3451 3400 Malaysia 603 9131 0918 Mexico 001 800 010 0793 Netherlands 31 0 348 433 466 New Zealand 0800 553 322 Nor...

Page 4: ...tion of the product owner s abuse misuse or negligent acts and power failure or surges fire flood accident actions of third parties or other events outside reasonable control Copyright Under the copyr...

Page 5: ...ons not expressly approved by NI could void the user s authority to operate the equipment under the FCC Rules Class A Federal Communications Commission This equipment has been tested and found to comp...

Page 6: ...sis Description 1 6 SCXI Subsystem Backplane Overview 1 9 PXI Subsystem Backplane Overview 1 9 Interoperability with CompactPCI 1 9 System Controller Slot 1 10 Peripheral Slots 1 10 Star Trigger Slot...

Page 7: ...ior Cleaning 3 1 Exterior Cleaning 3 2 PXI Subsystem Fan Filter Cleaning 3 2 SCXI Subsystem Fan Filter Cleaning 3 2 Resetting the AC Main Circuit Breaker 3 3 SCXI Subsystem Fuse Replacement 3 4 Power...

Page 8: ...ensions Top View Shown A 9 Tables Table 1 1 Power Cables 1 2 Table 1 2 PXI 1010 Chassis Front View Items 1 6 Table 1 3 PXI 1010 Chassis Rear View Items 1 8 Table 2 1 PXI 1010 Voltage Selection and Fus...

Page 9: ...information about the DAQ hardware that plugs into or is connected to your computer Use these documents for hardware installation and configuration instructions specification information about your DA...

Page 10: ...to a key concept This font also denotes text that is a placeholder for a word or value that you must supply monospace Text in this font denotes text or characters that you should enter from the keyboa...

Page 11: ...d helpful as you read this manual Your computer technical reference manual If you are designing your own module the following National Instruments specification available by request describes the phys...

Page 12: ...modules The four SCXI slots integrate signal conditioning modules into your PXI system These modules provide analog and digital input conditioning isolation and other functions The PXI 1010 is ideal f...

Page 13: ...visible damage to the metal work Check to make sure all handles hardware and switches are undamaged Inspect the inner chassis for any possible damage debris or detached components If damage appears t...

Page 14: ...l Instruments DAQ hardware is included with LabVIEW The LabVIEW Data Acquisition VI Library is functionally equivalent to NI DAQ software LabWindows CVI features interactive graphics state of the art...

Page 15: ...addresses many of the complex issues between the computer and the plug in board such as programming interrupts and DMA controllers NI DAQ maintains a consistent software interface among its different...

Page 16: ...ules and accessories for isolating amplifying exciting and multiplexing signals for relays and analog output PXI modules and accessories for multifunction I O and instrumentation For more specific inf...

Page 17: ...ripheral Slots 15 PXI Star Trigger Slot 16 PXI System Controller Slot 17 PXI Controller Expansion Slots Table 1 2 PXI 1010 Chassis Front View Items Item Description PXI module guides Guide PXI modules...

Page 18: ...us connections to SCXI modules SCXI Slot 0 power supply Contains the power supply and control circuitry for the SCXI subsystem PXI SCXI Control slot Device in this slot can control SCXI subsystem PXI...

Page 19: ...is Rear connector space For module space connector mounting brackets or adapter boards Power entry module IEC receptacle for power input voltage selection wheel and fuse Voltage selection wheel Config...

Page 20: ...r container You can use the chassis this way in addition to or instead of one of the first three methods PXI Subsystem Backplane Overview The following sections describe the PXI subsystem backplane In...

Page 21: ...ipheral slot refer to Figure 1 4 and is intended for modules with Star Trigger functionality to provide individual triggers to all other peripherals However if you do not require advanced trigger func...

Page 22: ...es in a variety of ways For example you can use triggers to synchronize the operation of several different PXI peripheral modules In other applications one module can control carefully timed sequences...

Page 23: ...signal to each peripheral slot You can use this common reference clock signal to synchronize multiple modules in a measurement or control system You can drive PXI_CLK10 from an external source throug...

Page 24: ...es along both sides of the chassis have adequate ventilation Keep other equipment a minimum of 3 0 in 76 2 mm away from the air inlets and outlets Install your chassis for easy access to the rear pane...

Page 25: ...eel and fuse should be correct for operation in your geographical region the 0x extension of the kit part number corresponds to your geographical region Check the voltage on the voltage selection whee...

Page 26: ...crewdriver pry the door to the selection wheel open from the top 4 Remove the selection wheel 5 Rotate the wheel for the appropriate voltage and reinsert it into the power entry module 6 Close the doo...

Page 27: ...selection wheel in the power entry module is set for your outlet line voltage Refer to the Line Voltage Selection and Fuse Values section earlier in this chapter if necessary 5 Insert the female end o...

Page 28: ...aking sure that the injector ejector handle is pushed down as shown in Figure 2 3 3 When you begin to feel resistance push up on the injector ejector handle to inject the module into the frame 4 Secur...

Page 29: ...n in Figure 1 2 Front View of the PXI 1010 Chassis 2 Remove the front filler panel of an empty SCXI slot 3 Insert the module into the module guide rails and slide the module to the back of the chassis...

Page 30: ...threaded strip holes Rear Filler Panel Installation Rear filler panels are half the height of front filler panels To install SCXI rear filler panels complete the following steps 1 Place the filler pan...

Page 31: ...ents within the chassis are susceptible to static discharge damage Service the chassis only in a static free environment Observe standard handling precautions for static sensitive devices while servic...

Page 32: ...or blow air through them Rinse the filters with water and allow them to dry before reinstalling them onto the chassis SCXI Subsystem Fan Filter Cleaning The fan on the rear of the SCXI subsystem cool...

Page 33: ...e PXI 1010 Chassis as you complete the following steps 1 Turn off the PXI power switch 2 Depress the circuit breaker to reset it 3 Turn on the PXI power switch If the circuit breaker trips again compl...

Page 34: ...r 7 Reinsert the fuse drawer in the right side slot with the arrow pointing to the right 8 Close the door 9 Reinsert the power cable SCXI Subsystem Backplane Fuse Replacement and Check In addition to...

Page 35: ...cm 0 25 in 5 Using the needle nose pliers insert the fuse into the socket holes 6 Repeat if necessary for the other fuse 7 Check the fan filter and if it is dirty clean it as described in the SCXI Su...

Page 36: ...to 240 VAC Operating voltage range1 90 to 264 VAC Input frequency 50 60 Hz Operating frequency range1 47 to 63 Hz Max current 120 VAC 10 8 A 100 VAC 10 8 A 240 VAC 10 4 A 220 VAC 10 4 A Over current p...

Page 37: ...e CompactPCI PXI backplane PXI Subsystem DC Output DC current capacity Load regulation Maximum ripple and noise 20 MHz bandwidth Voltage IMP Steady state current 3 3 V 30 A 5 V 20 A 12 V 4 A 12 V 2 A...

Page 38: ...aracteristics DC Output V output Tolerance limits include peaks 18 5 to 25 V Ripple peak to peak 1 5 V Max load 680 mA V output Tolerance limits include peaks 18 5 to 25 V Ripple peak to peak 1 5 V Ma...

Page 39: ...sis Power supply cooling exhaust Sides of chassis Module cooling fan MTBF 30 000 h Environment Temperature IEC 60068 2 1 2 0 to 50 C functional 20 to 70 C storage Humidity IEC 60068 2 3 10 to 90 nonco...

Page 40: ...y use EN 61010 1 IEC 61010 1 UL 3111 1 UL 61010B 1 CAN CSA C22 2 No 1010 1 Note For UL and other safety certifications refer to the product label or visit ni com hardref nsf search by model number or...

Page 41: ...ink in the Certification column Backplane Size 3U sized one system slot with three system expansion slots and seven peripheral slots Compliant with IEEE 1101 10 mechanical packaging PXI Specification...

Page 42: ...r Slot Input frequency 10 MHz 100 ppm or better Input amplitude Slot 2 5 V or 3 3 V 10 MHz TTL signal Maximum jitter introduced by backplane circuitry 1 ps RMS in 10 Hz to 1 MHz range Mechanical Overa...

Page 43: ...Shown Figure A 2 PXI 1010 Dimensions Rear View Shown 1 PXI Subsystem 8 slots 2 SCXI Subsystem 4 slots 9 820 in 243 1 mm 6 970 in 177 04 mm 17 250 in 438 15 mm 1 2 9 825 in 249 1 mm 17 250 in 438 56 mm...

Page 44: ...ser Manual Figure A 3 PXI 1010 Dimensions Top View Shown 17 250 in 438 15 mm 9 825 in 249 56 mm 9 570 in 243 1 mm 1 212 in 30 78 mm 17 600 in 447 07 mm 16 380 in 416 04 mm 16 241 in 412 52 mm Artisan...

Page 45: ...onnector pin assignments for the System Controller slot Table B 3 shows the P1 J1 connector pin assignments for the Star Trigger slot Table B 4 shows the P2 J2 connector pin assignments for the Star T...

Page 46: ...3V PAR C BE 1 GND 17 GND 3 3V SDONE SBO GND PERR GND 16 GND DEVSEL GND V I O STOP LOCK GND 15 GND 3 3V FRAME IRDY GND TRDY GND 12 14 Key Area 11 GND AD 18 AD 17 AD 16 GND C BE 2 GND 10 GND AD 21 GND...

Page 47: ...TRIG7 GND 15 GND PXI_BRSVA15 GND FAL REQ5 GNT5 GND 14 GND AD 35 AD 34 AD 33 GND AD 32 GND 13 GND AD 38 GND V I O AD 37 AD 36 GND 12 GND AD 42 AD 41 AD 40 GND AD 39 GND 11 GND AD 45 GND V I O AD 44 AD...

Page 48: ...PAR C BE 1 GND 17 GND 3 3V SDONE SBO GND PERR GND 16 GND DEVSEL GND V I O STOP LOCK GND 15 GND 3 3V FRAME IRDY GND TRDY GND 12 14 Key Area 11 GND AD 18 AD 17 AD 16 GND C BE 2 GND 10 GND AD 21 GND 3 3...

Page 49: ...I_TRIG7 GND 15 GND PXI_BRSVA15 GND FAL PXI_STAR6 PXI_LBR6 GND 14 GND AD 35 AD 34 AD 33 GND AD 32 GND 13 GND AD 38 GND V I O AD 37 AD 36 GND 12 GND AD 42 AD 41 AD 40 GND AD 39 GND 11 GND AD 45 GND V I...

Page 50: ...PAR C BE 1 GND 17 GND 3 3V SDONE SBO GND PERR GND 16 GND DEVSEL GND V I O STOP LOCK GND 15 GND 3 3V FRAME IRDY GND TRDY GND 12 14 Key Area 11 GND AD 18 AD 17 AD 16 GND C BE 2 GND 10 GND AD 21 GND 3 3V...

Page 51: ...TRIG7 GND 15 GND PXI_BRSVA15 GND FAL PXI_LBL6 PXI_LBR6 GND 14 GND AD 35 AD 34 AD 33 GND AD 32 GND 13 GND AD 38 GND V I O AD 37 AD 36 GND 12 GND AD 42 AD 41 AD 40 GND AD 39 GND 11 GND AD 45 GND V I O A...

Page 52: ...an answer Training and Certification Visit ni com training for self paced training eLearning virtual classrooms interactive CDs and Certification program information You also can register for instruct...

Page 53: ...2 ni com the Worldwide Offices section of ni com niglobal to access the branch office Web sites which provide up to date contact information support phone numbers email addresses and current events Ar...

Page 54: ...unpredictable time with respect to the execution of a program 2 Hardware a property of an event that occurs at an arbitrary time without synchronization to a reference clock B backplane An assembly t...

Page 55: ...D daisy chain A method of propagating signals along a bus in which the devices are prioritized on the basis of their position on the bus DC Direct current E EMC Electromagnetic Compatibility F FCC Fed...

Page 56: ...stitute of Electrical and Electronics Engineers IMP Mainframe peak current in Inches L lb Pounds local bus A group of signals that connect two peripheral slots M m Meters MTBF Mean time between failur...

Page 57: ...I module that plugs into a PXI peripheral slot PXI peripheral slot Holds PXI peripheral modules PXI subsystem The PXI 1010 chassis section that uses PXI modules R RMS Root mean square A method used to...

Page 58: ...s system controller functions including clock sourcing and arbitration for data transfers across the backplane Installing such a device into any other slot can damage the device the VXIbus backplane o...

Page 59: ...and fuse values 2 2 fuse part numbers table 2 3 voltage selection and fuse ratings by region table 2 3 SCXI subsystem address selection 2 1 conventions used in the manual x D Declaration of Conformity...

Page 60: ...odule 2 6 unpacking PXI 1010 chassis 1 2 instrument drivers NI resources C 1 K key features of the PXI 1010 chassis 1 5 KnowledgeBase C 1 L LabVIEW software 1 3 LabWindows CVI software 1 3 line voltag...

Page 61: ...n figure 2 6 procedure 2 4 PXI or CompactPCI modules figure 2 5 interoperability with CompactPCI 1 9 local bus 1 10 peripheral slots 1 10 SCXI connection 1 11 Star Trigger and local bus routing figure...

Page 62: ...magnetic compatibility A 5 environmental A 4 mechanical A 7 safety A 5 Star Trigger slot description 1 10 figure 1 6 P1 J1 connector pin assignments table B 4 P2 J2 connector pin assignments table B 5...