CALIBRATION PROCEDURE
NI PXIe-4463
DSA Analog Output
This document contains the verification and adjustment procedures for the National Instruments
PXIe-4463 with either BNC or mini-XLR connectors. For more information about calibration
solutions, visit
Contents
Software.................................................................................................................................... 2
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 5
Initial Setup............................................................................................................................... 6
AO Offset Accuracy Verification............................................................................. 7
AO Gain Accuracy Verification ............................................................................... 9
AO Flatness Verification .......................................................................................... 11
AO Interchannel Gain Mismatch Verification ......................................................... 12
AO Noise Verification.............................................................................................. 13
AO Current Drive Verification................................................................................. 15
AO Offset and Gain Adjustment ...................................................................................... 18
Timebase Adjustment ....................................................................................................... 19