b)
Input the NI 4113 measurements as the
reported outputs
.
c)
Input the specified level range as the
range
.
9.
Update the output calibration constants by configuring and calling the niDCPower Cal
Adjust Current Limit VI or
niDCPower_CalAdjustCurrentLimit
function.
a)
Input the calculated shunt current measurements as the
measured outputs
.
b)
Input the test points as the
requested outputs
.
c)
Input the specified level range as the
range
.
10. If more than one level range is specified, repeat the previous steps using the values
specified in each level range.
Related Information
Closing the Adjustment Session
Close the session and commit the new constants to hardware by calling the niDCPower
Close External Calibration VI or
niDCPower_CloseExtCal
function and specifying
Commit as the
calibration close action
.
Reverification
Repeat the
Verification
section to determine the as-left status of the device.
Note
If any test fails reverification after performing an adjustment, verify that you
have met the
Test Conditions
before returning your device to NI. Refer to the
Worldwide Support and Services
section for information about support resources or
service requests.
Related Information
Worldwide Support and Services
The NI website is your complete resource for technical support. At
access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.
Visit
for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit
to register your NI product. Product registration facilitates technical
support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the
European Communities using the manufacturer’s declaration of conformity. This system
affords the user protection for electromagnetic compatibility (EMC) and product safety. You
NI PXIe-4113 Calibration Procedure
|