Table 1.

 Required Equipment for Calibration

Required

Equipment

Recommended

Model(s)

Parameter

Measured

Minimum

Specifications

 

 

 

 

One digital
multimeter
(DMM), optional
second DMM 

1

NI PXI-4071

All parameters

Voltage: <±200 ppm
accuracy and <30 µV
resolution

Programmable
electronic load

Agilent N3302A or
Sorensen
SLM-60-30-150

Load regulation

Constant current and
constant voltage
modes: sink at least
6 A at 10 V

100 mΩ precision
current shunt

Guildline 9230 A-30 or
Ohm Labs CS10

Current output
and measurement,
voltage and
current load
regulation

±100 ppm tolerance,
±50 ppm stability,
±5 ppm/C temperature
coefficient. Minimum
current 10 A.

Two 50 Ω resistors Vishay

PTF5650R000BZEK

Remote sense
output

0.1% tolerance, 1/8 W

1 kΩ resistor

Vishay
PTF651K0000BYEK

Remote sense
output

0.1% tolerance, 1/4 W

Test Conditions

The following setup and environmental conditions are required to ensure the NI 4113 meets
published specifications:

Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that
can affect measurements.

Verify that all connections to the device, including front panel connections and screws,
are secure.

Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters (if present) are
clean, and that the empty slots contain slot blockers and filler panels. For more
information about cooling, refer to the 

Maintain Forced-Air Cooling Note to Users

document available at 

ni.com/manuals

.

1

If you make measurements using only one DMM, sequentially set up the DMM connections as
specified by the procedure steps.

NI PXIe-4113 Calibration Procedure

  | 

© National Instruments

  | 

3

Summary of Contents for NI PXIe-4113

Page 1: ...ent 7 Load Regulation 9 Remote Sense 12 Adjustment 13 Adjusted Specifications 13 Initiating the Adjustment Session 14 Connecting and Configuring Equipment for Voltage Adjustment 14 Adjusting Voltage O...

Page 2: ...rform the calibration procedure NI PXIe 4112 4113 Getting Started Guide NI DC Power Supplies and SMUs Help NI PXIe 4113 Specifications NI DCPower Readme LabVIEW Help Visit ni com manuals for the lates...

Page 3: ...ce 1 8 W 1 k resistor Vishay PTF651K0000BYEK Remote sense output 0 1 tolerance 1 4 W Test Conditions The following setup and environmental conditions are required to ensure the NI 4113 meets published...

Page 4: ...ower line cycles Do not use the NI DCPower Soft Front Panel SFP to request test points for any adjustment functions because you cannot set aperture time using the SFP Ensure that properties or attribu...

Page 5: ...hether the device is operating within the published specifications prior to adjustment 3 Adjustment Adjust the calibration constants of the NI 4113 4 Reverification Repeat the Verification procedure t...

Page 6: ...he NI 4113 2 Set the level on the NI 4113 to the first specified test point 3 Compare a DMM voltage measurement to the voltage output test limits a Take a voltage measurement using the DMM b Calculate...

Page 7: ...ts DMM Measured Voltage DMM Measured Voltage of Voltage Offset c Verify the NI 4113 measurement falls within the test limits 4 Repeat the previous two steps for each test point specified in the level...

Page 8: ...late the lower and upper current output test limits using the following formula Current Output Test Limits Test Point Test Point of Current Offset 5 Verify that the recorded DMM value falls within the...

Page 9: ...fied in the level range 9 If more than one level range is specified repeat the previous steps using the values specified in each level range Load Regulation Connecting and Configuring Equipment for Vo...

Page 10: ...across the output terminals of the NI 4113 using DMM1 5 Record the voltage from the previous step as V1 6 Calculate the current through the shunt by completing the following steps a Take a voltage me...

Page 11: ...g current output and measurement Refer to the following table as you complete the following steps Table 7 Current Load Regulation Verification Level Range Limit Range and Limit Test Point Load1 Load2...

Page 12: ...ion test limits 12 If more than one level range is specified repeat the previous steps using the values specified in each level range Remote Sense Connecting and Configuring Equipment for Voltage Remo...

Page 13: ...th DMM1 from the negative terminal of the device to the negative side of Load2 Record the measurement as Lead Drop 4 Verify the Lead Drop measurement is greater than or equal to the minimum lead drop...

Page 14: ...lete each adjustment procedure Related Information Closing the Adjustment Session on page 17 Connecting and Configuring Equipment for Voltage Adjustment 1 Make the necessary connections for this proce...

Page 15: ...tion a Input the DMM measurements as the measured outputs b Input the NI 4113 measurements as the reported outputs c Input the specified level range as the range 7 Update the output calibration consta...

Page 16: ...the level on the NI 4113 to the first specified test point 3 Calculate the current through the shunt by completing the following steps a Take a voltage measurement across the shunt using the DMM b Div...

Page 17: ...status of the device Note If any test fails reverification after performing an adjustment verify that you have met the Test Conditions before returning your device to NI Refer to the Worldwide Suppor...

Page 18: ...rein are trademarks or trade names of their respective companies For patents covering NI products technology refer to the appropriate location Help Patents in your software the patents txt file on you...

Reviews: