3

NI PXI-665x Calibration Procedure

Calibration Interval

The measurement requirements of the application determine how often you 
should calibrate the device to ensure its accuracy. National Instruments 
recommends that you calibrate the NI PXI-6653 and NI PXI-6652 

yearly

You may want to shorten this interval based on the application demands.

Password

The default password for password-protected operations is 

NI

.

Test Equipment

Calibrating the NI PXI-665

x

 requires the following equipment.

Test Conditions

Follow these guidelines to optimize connections and test conditions during 
the calibration procedure:

Install the NI PXI-665

x

 in Slot 2 of the PXI chassis. The calibration 

procedure requires features of PXI that are accessible only in slot 2. 
Use of an NI PXI-1044 or NI PXI-1045 chassis is recommended to 
reduce error associated in variance in the 5 V chassis supply.

Maintain a temperature of approximately 25 °C.

Keep relative humidity below 80 percent.

Use shielded copper wire for all cable connections to the device.

Allow a warm up time of at least 15 minutes for the NI PXI-6652 and 
three hours for the NI PXI-6653 to ensure the measurement circuitry is 
at a stable operating temperature.

Keep PXI chassis filters clean and fan speed set to High.

Equipment

Recommended Model

Requirements

10 MHz clock source

Symmetricom/Datum 8040

Accurate to within 
0.75 parts-per-billion (ppb) for the 
NI PXI-6653, or 150 ppb for the 
NI PXI-6652.

BNC-SMB cable

SMB-SMB cable

Summary of Contents for NI PXI-665 Series

Page 1: ...g the Clock Source 4 Measuring the Frequency 6 Adjustment 10 Reverification 13 Where to Go for Support 14 Conventions The following conventions apply to this document The symbol leads you through nest...

Page 2: ...ns Software This calibration procedure requires NI Sync and NI VISA NI Sync and NI VISA support a number of application development environments ADEs and programming languages including LabVIEW LabWin...

Page 3: ...the NI PXI 665x in Slot 2 of the PXI chassis The calibration procedure requires features of PXI that are accessible only in slot 2 Use of an NI PXI 1044 or NI PXI 1045 chassis is recommended to reduc...

Page 4: ...e driver and application software is installed on the host computer 5 Configure the hardware properly with Measurement Automation Explorer MAX Refer to the NI PXI 665x Installation Guide for details a...

Page 5: ...tion Call Call niSync_init with the following parameters resourceName Dev1 idQuery NULL resetDevice VI_TRUE vi SessionHandle LabVIEW Block Diagram NI SYNC C Function Call Call niSync_SetAttributeViBoo...

Page 6: ...nc Initialize VI to set up a handle for the device b Set a writable NI Sync property node to pass TRUE to the ClkOut Gain Enable attribute LabVIEW Block Diagram NI SYNC C Function Call Call niSync_clo...

Page 7: ...on must be made sufficiently long Refer to Table 1 for the measurement duration to use LabVIEW Block Diagram NI SYNC C Function Call Call niSync_ConnectClkTerminals with the following parameters vi Se...

Page 8: ...agram NI SYNC C Function Call Call niSync_init with the following parameters resourceName Dev1 idQuery NULL resetDevice VI_FALSE vi SessionHandle LabVIEW Block Diagram NI SYNC C Function Call Call niS...

Page 9: ...r period following a calibration Table 2 One Year Test Limits Device Specified Accuracy Aging Acceptable Frequency Range As Found Low Limit High Limit NI PXI 6652 TCXO 3 5 ppm 9 999 965 Hz 10 000 035...

Page 10: ...plete steps a c to set the oscillator control voltage The range of acceptable voltage values is 0 0 V to 4 095 V with frequency increasing as voltage increases Use a control voltage of 2 0475 V which...

Page 11: ...quency section To obtain an accurate measurement for calibration the measurement duration must be made sufficiently long Refer to Table 1 Measurement Duration for the measurement duration to use Compl...

Page 12: ...Ranges 5 Commit the calibration values to the Calibration EEPROM using the following procedure a Call niSync Initialize External Calibration VI to initialize the process Note NI is the default user p...

Page 13: ...on section Re verifying after making the adjustments ensures that the NI PXI 665x is operating within its test limits LabVIEW Block Diagram NI SYNC C Function Call Call niSync_InitExtCal with the foll...

Page 14: ...s National Instruments corporate headquarters is located at 11500 North Mopac Expressway Austin Texas 78759 3504 National Instruments also has offices located around the world to help address your sup...

Reviews: